Toe blood pressure measurements in patients suspected of leg ischaemia: a new laser Doppler device compared with photoplethysmography.

机译:怀疑患有腿部缺血的患者的脚趾血压测量:与光电容积描记术相比,一种新型的激光多普勒仪。

获取原文
获取原文并翻译 | 示例
       

摘要

Objectives. To study the usefulness of a new laser Doppler (LD) device as compared with photoplethysmography (PPG) to measure toe systolic blood pressures (TBP) in patients suspected of arterial insufficiency of the legs. Design. Prospective comparative clinical study. Materials and Methods. Seventy patients (121 legs) referred to our vascular laboratory for TBP measurements were investigated using a standard PPG and a new LD instrument, incorporating a cuff inflator and probe heater. Results. Mean toe pressures with the PPG, unheated LD, and heated LD (37 degrees C) were 68.1, 67.2, and 75.5 mmHg, respectively. TBP with heated LD was systematically higher than with PPG (7.7 mmHg; p<0.001). Intra-class correlations between PPG vs. unheated LD and PPG vs. heated LD were good: 0.91 and 0.93, respectively. However, a considerable variation existed between PPG and LD (95% confidence interval: 45 mmHg), which were unrelated to the value of the TBP. Kappa values of agreement between PPG and unheated and heated LD to detect critical ischaemia (TBP below 30 mmHg) were good (kappa=0.77 and 0.63, respectively). Conclusions. The new LD device appears a good alternative to standard PPG measurements to assess TBP, despite a rather wide variation.
机译:目标。为了研究怀疑是腿部动脉供血不足的患者的新型激光多普勒(LD)装置与光电容积描记术(PPG)相比,其测量脚趾收缩压(TBP)的有用性。设计。前瞻性比较临床研究。材料和方法。使用标准PPG和新的LD仪器(包括袖套充气机和探头加热器)对70名患者(121腿)转诊到我们的血管实验室进行了TBP测量。结果。 PPG,未加热的LD和加热的LD(37摄氏度)的平均脚趾压力分别为68.1、67.2和75.5 mmHg。 LD加热时的TBP系统性高于PPG(7.7 mmHg; p <0.001)。 PPG对未加热的LD和PPG对加热的LD之间的类内相关性很好:分别为0.91和0.93。但是,PPG和LD之间存在相当大的差异(95%置信区间:45 mmHg),与TBP的值无关。 PPG与未加热和加热的LD之间检测严重缺血(TBP低于30 mmHg)的协议Kappa值良好(kappa分别为0.77和0.63)。结论尽管变化很大,但新的LD器件似乎可以很好地替代标准PPG测量来评估TBP。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号