Frame dislocation of body middle rings in endovascular stent tube grafts.

机译:血管内支架管移植物中身体中环的框架脱位。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVES: To understand the cause, and propose a mechanism for frame dislocation in endovascular grafts. MATERIALS AND METHODS: Five tube grafts were explanted due to secondary distal leakage 15-21 months after operation. One bifurcated graft was removed during emergency operation after aortic rupture caused by secondary leakage. A second bifurcated graft was harvested from a patient with thrombotic occlusion of one limb, who died after transurethral prostatic resection. The inside of the grafts were examined endoscopically. The stent was inspected after removal of the fabric, broken ligatures were counted and examined by scanning electron microscopy. The fabric strength was tested by probe puncture. RESULTS: We found 17-44% of the stent ligatures of the body middle rings to be loose. The knots were intact. Degradation of the polyester textile was not observed. CONCLUSIONS: Continuous movements in the grafted aorta and blood pressure impose permanent stress to the stent frame and the polyester fabric resulting in morphological changes in the body middle ring of grafts. The clinical implications of the suture breakages are unknown although they may be related to distal secondary leakage in tube grafts.
机译:目的:了解原因,并提出血管内移植物中框架脱位的机制。材料与方法:由于术后15-21个月继发远端漏血,将五管移植物植入。由二次渗漏引起的主动脉破裂后,在紧急手术期间取出了一根分叉的移植物。从一名患肢的血栓闭塞的患者中收获第二个分叉的移植物,该患者在经尿道前列腺切除术后死亡。内窥镜检查移植物的内部。除去织物后检查支架,计数断裂的绑扎线并通过扫描电子显微镜检查。通过探针刺穿测试织物强度。结果:我们发现身体中环的支架结扎中有17-44%松动。结完好无损。未观察到聚酯纺织品的降解。结论:连续的主动脉运动和血压升高会给支架框架和聚酯纤维施加永久性应力,从而导致人体中部环的形态发生变化。缝线破裂的临床意义尚不清楚,尽管它们可能与管移植物的远端继发性渗漏有关。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号