A method for direct potentiometric determination of lead(II) ion in native, untreated environmental samples of groundwater is reported. In a filtered sample, when applying a solid-state (PbS/Ag_2S) ion-selective electrode in tuned galvanostatic polarization (TGP) mode, a concentration of 17.2±1.6μgdm~(-3) was measured. The potentiometric results are compared with those obtained using inductively coupled plasma mass spectrometry (ICP-MS), 18.0±0.4μgdm~(-3) and differential pulse anodic stripping voltammetry (DPASV), 24.3±0.5μgdm~(-3). Owing to limitations of other methods, the analysis of Pb~(2+) in the untreated and non-filtered samples was only possible with the ion-selective electrode solid-state electrode. The described method can be an attractive alternative to existing methods of Pb~(2+) determination that are used to monitor groundwater pollution.
机译:报道了一种直接电位法测定天然未经处理的地下水环境样品中铅(II)离子的方法。在过滤后的样品中,当以调谐恒电流极化(TGP)模式应用固态(PbS / Ag_2S)离子选择电极时,测得的浓度为17.2±1.6μgdm〜(-3)。将电位结果与使用电感耦合等离子体质谱(ICP-MS),18.0±0.4μgdm〜(-3)和差分脉冲阳极溶出伏安法(DPASV),24.3±0.5μgdm〜(-3)获得的结果进行比较。由于其他方法的局限性,只有使用离子选择电极固态电极才能对未处理和未过滤样品中的Pb〜(2+)进行分析。所描述的方法可以替代现有的监测地下水污染的Pb〜(2+)测定方法。