首页> 外文期刊>Electroanalysis >A Sensitive Electrochemical Sensor Based on Solution Polymerized Molecularly Imprinted Polymers for Procaine Detection
【24h】

A Sensitive Electrochemical Sensor Based on Solution Polymerized Molecularly Imprinted Polymers for Procaine Detection

机译:基于溶液聚合分子印迹聚合物的敏感电化学传感器用于普鲁卡因的检测

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

A sensitive and selective method for the determination of procaine hydrochloride using molecularly imprinted polymers (MIPs) modified glassy carbon electrodes was developed. The MIPs were prepared by solution polymerization using procaine hydrochloride as the template molecules and acrylic acid (AA) and vinyltriethoxysilane (WD-20) as the functional monomer and crosslinking agent, respectively. A film was formed on the surface of the glassy carbon electrodes and later cross-linked with ethanol as solvent. Next, these electrodes were employed to detect procaine hydrochloride by differential pulse voltammetry (DPV). Under the optimized conditions, good linearity (correlation coefficient of 0.9986) was observed between the oxidation peak current and the concentration of procaine hydrochloride in the range of 4.0 x 10(-8) to 2.4 x 10(-5) M in a pH 7.0 0.04 M phosphate buffer solution, and the detection limit (S/N=3) was 1.02 x 10(-8) M. In addition, the stability and reproducibility of the sensors were satisfactory. Moreover, the concentration of procaine hydrochloride in human blood serum samples was detected, and its recoveries ranged from 97.5% to 106.4% with RSD less than 2.15%. These results suggest that the prepared molecularly imprinted electrochemical sensors can be used for the determination of procaine in clinical studies.
机译:建立了一种灵敏且选择性的测定分子印迹聚合物(MIPs)修饰的玻碳电极测定盐酸普鲁卡因的方法。通过分别使用盐酸普鲁卡因作为模板分子和丙烯酸(AA)和乙烯基三乙氧基硅烷(WD-20)作为功能单体和交联剂的溶液聚合反应制备MIP。在玻璃碳电极的表面上形成膜,然后用乙醇作为溶剂交联。接下来,这些电极用于通过差分脉冲伏安法(DPV)检测盐酸普鲁卡因。在最佳条件下,在pH 7.0时,氧化峰电流与盐酸普鲁卡因浓度之间的线性关系良好(相关系数为0.9986),范围为4.0 x 10(-8)至2.4 x 10(-5)M 0.04 M磷酸盐缓冲溶液,检出限(S / N = 3)为1.02 x 10(-8)M。此外,传感器的稳定性和重现性令人满意。此外,检测到人血清样品中盐酸普鲁卡因的浓度,其回收率在97.5%至106.4%之间,RSD小于2.15%。这些结果表明,所制备的分子印迹电化学传感器可用于临床研究中普鲁卡因的测定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号