Determination of clozapine by capillary zone electrophoresis following end-column amperometric detection with simplified capillary/electrode alignment

机译:末端毛细管安培检测后通过毛细管区带电泳测定氯氮平,简化了毛细管/电极对准

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Capillary zone electrophoresis was employed for the determination of clozapine using an end-column amperometric detection at a carbon fiber array microdisk electrode with simplified capillary/electrode alignment. The optimum conditions of separation and detection are: Britton-Robinson butter, pH 2.0 (1.3 x 10(-2) mol/L total concentration of acids, 3.2 x 10(-3) mol/L NaOH), 15 kV for separation voltage, 5 kV and 10 s for injection voltage and injection time, respectively. The limit of detection is 4.2 x 10(-7) mol/L or 1.2 fmole (signal to noise, S/N = 2). The relative standard deviation is 1.4% for the migration time and 2.5% for the electrophoretic peak current. The method was applied to the determination of clozapine in human blood. The recovery of the method is between 94-104%. [References: 28]
机译:毛细管区带电泳用于在碳纤维阵列微盘电极上以简化的毛细管/电极对齐方式进行端柱安培检测来测定氯氮平。分离和检测的最佳条件是:Britton-Robinson黄油,pH 2.0(酸的总浓度为1.3 x 10(-2)mol / L,NaOH为3.2 x 10(-3)mol / L),分离电压为15 kV ,分别为5 kV和10 s的注入电压和注入时间。检测极限为4.2 x 10(-7)mol / L或1.2 fmole(信噪比,S / N = 2)。迁移时间的相对标准偏差为1.4%,电泳峰电流的相对标准偏差为2.5%。该方法用于人血中氯氮平的测定。该方法的回收率在94-104%之间。 [参考:28]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号