...
首页> 外文期刊>Investigative radiology >Preclinical evaluation of MR attenuation correction versus CT attenuation correction on a sequential whole-body MR/PET scanner
【24h】

Preclinical evaluation of MR attenuation correction versus CT attenuation correction on a sequential whole-body MR/PET scanner

机译:在连续全身MR / PET扫描仪上进行MR衰减校正与CT衰减校正的临床前评估

获取原文
获取原文并翻译 | 示例

摘要

OBJECTIVES: The application of attenuation correction for combined magnetic resonance/positron emission tomography (MR/PET) systems is still a major challenge for accurate quantitative PET. Computed tomographic attenuation correction (CTAC) is the current clinical standard for PET/computed tomographic (CT) scans. Magnetic resonance, unlike CT, has no direct information about photon attenuation but, rather, proton densities. On combined MR/PET scanners, MR-based attenuation correction (MRAC) consists of assigning empirical attenuation coefficients to MR signal intensities. The objective of the current study was to evaluate the MRAC implemented on the combined MR/PET scanner versus the CTAC with the same PET data in an animal model. MATERIALS AND METHODS: Acquisition was performed using a clinically approved sequential MR/PET scanner (Philips Ingenuity TF). Computed tomographic and MR/PET images of 20 New Zealand White rabbits were retrospectively analyzed. The animals were positioned on a customized animal bed to avoid movement between the CT and MR/PET scanners. Positron emission tomographic images from both methods (MRAC and CTAC) were generated. Voxel-by-voxel and region-of-interest (ROI) analyses were performed to determine differences in standardized uptake values (SUV). Regions of interest were drawn on the coregistered CT images for the aorta, liver, kidney, spine, and soft tissue (muscle) and superimposed on the PET images. RESULTS: The voxel-by-voxel comparison of PET showed excellent correlation between MRAC and CTAC SUV values (R = 0.99; P < 0.0001). The mean of the difference of SUVs between all respective MRAC and CTAC voxels was -0.94% (absolute difference [AD] ± SD, -0.06 ± 0.30), confirming slight underestimation of MRAC. The ROI-based comparison similarly showed that MRAC SUV values were underestimated compared with CTAC SUV values. The mean difference between MRAC and CTAC for all ROIs was 10.8% (AD, -0.08 ± 0.06; R = 0.99; P < 0.0001) and -9.7% (AD, -0.15 ± 0.12; R = 0.99; P < 0.0001) for the SUV mean (SUVmean) and the SUV maximum (SUVmax), respectively. The highest differences were found in the spine (SUVmean -26.1% [-0.11]) and areas close to large bones such as the back muscles (SUVmean, -16.8% [-0.04]). CONCLUSIONS: In this study, we have compared MRAC and CTAC methods for PET attenuation correction in an animal model. We have confirmed that the MRAC method implemented on a sequential MR/PET scanner underestimates PET values by less than 10% in most regions, except the areas containing or close to large bone structures such as the spine or the back muscles. Bone segmentation is therefore suggested to be included in the MR attenuation map to minimize the quantification error of MRAC methods compared with the clinical standard CTAC. Further clinical studies need to be carried out to validate the clinical use of MRAC.
机译:目标:衰减校正在磁共振/正电子发射断层扫描(MR / PET)系统中的应用仍然是精确定量PET的主要挑战。计算机断层扫描衰减校正(CTAC)是PET /计算机断层扫描(CT)扫描的当前临床标准。磁共振与CT不同,它不具有有关光子衰减的直接信息,而具有质子密度。在组合的MR / PET扫描仪上,基于MR的衰减校正(MRAC)包括为MR信号强度分配经验衰减系数。本研究的目的是评估在动物模型中具有相同PET数据的MR / PET组合扫描仪与CTAC结合实施的MRAC。材料与方法:使用临床认可的顺序MR / PET扫描仪(Philips Ingenuity TF)进行采集。回顾性分析了20只新西兰白兔的计算机断层扫描和MR / PET图像。将动物放置在定制的动物床上,以避免在CT和MR / PET扫描仪之间移动。两种方法(MRAC和CTAC)产生的正电子发射断层图像。进行了逐像素和关注区域(ROI)分析,以确定标准摄取值(SUV)的差异。在主动脉,肝脏,肾脏,脊柱和软组织(肌肉)的共同注册CT图像上绘制感兴趣的区域,并将其叠加在PET图像上。结果:PET的逐像素比较显示MRAC与CTAC SUV值之间具有极好的相关性(R = 0.99; P <0.0001)。所有MRAC和CTAC体素之间SUV的差异平均值为-0.94%(绝对差异[AD]±SD,-0.06±0.30),证实了MRAC的低估。基于ROI的比较类似地显示,与CTAC SUV值相比,MRAC SUV值被低估了。对于所有ROI,MRAC和CTAC之间的平均差异为10.8%(AD,-0.08±0.06; R = 0.99; P <0.0001)和-9.7%(AD,-0.15±0.12; R = 0.99; P <0.0001) SUV平均数(SUVmean)和SUV最大数(SUVmax)。在脊柱(SUVmean为-26.1%[-0.11])和靠近大骨头(如背部肌肉)的区域(SUVmean为-16.8%[-0.04]),差异最大。结论:在这项研究中,我们比较了MRAC和CTAC方法对动物模型中PET衰减的校正。我们已经确认,在连续的MR / PET扫描仪上实施的MRAC方法在大多数区域(包含或接近大型骨骼结构(例如脊柱或背部肌肉)的区域)低估了PET值不到10%。因此,与临床标准CTAC相比,建议在MR衰减图中包括骨分割,以最大程度地减少MRAC方法的定量误差。需要进行进一步的临床研究以验证MRAC的临床用途。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号