Heat profiling of phacoemulsification tip using a thermal scanning camera

机译:使用热扫描相机对超声乳化尖端进行热分析

获取原文
获取原文并翻译 | 示例
           

摘要

An experimental study to measure the heat profile of the phacoemulsification (phaco) tip using standard continuous phaco and hyperpulse phaco with and without waveform power modulation in the Millennium Microsurgical System with Custom Control Software (CCS). The phaco tip was imaged in air using a thermal camera. The highest temperature was measured 15 s after application of phaco power. Continuous, hyperpulse and waveform power modulations of the Millennium Microsurgical System were used with different power settings (20, 50 and 100 %) and duty cycles (40, 60 and 90 %), with the irrigation turned on and off. Using continuous phaco with the irrigation on, the phaco tip temperature remains <28.0 C. With irrigation off, the temperature is higher compared to irrigation on but still remains <45.0 C. Comparing the temperatures for all three power modulations when irrigation is on, at each phaco power and duty cycle setting, the temperature of the phaco tip is highest with continuous phaco, followed by hyperpulse with rise time 1, then hyperpulse with rise time 2. When irrigation is off, the highest temperatures are recorded using the hyperpulse with rise time 2, followed by continuous phaco, then hyperpulse with rise time 1. Hyperpulse and waveform modulations reduce heat generation compared to the continuous mode when irrigation is turned on. Lower duty cycles and lower ultrasound power produce less heat at the phaco tip.
机译:在带有自定义控制软件(CCS)的Millennium显微外科系统中,使用标准连续超声和超脉冲超声在测量和不进行波形功率调制的情况下,对超声乳化(phaco)尖端的热量分布进行测量的实验研究。使用热像仪在空气中对超声镜尖端成像。施加phaco功率15 s后,测量最高温度。千年显微外科手术系统的连续,超脉冲和波形功率调制用于不同的功率设置(20%,50%和100%)和占空比(40%,60%和90%),并且打开和关闭冲洗功能。在连续灌溉的情况下使用连续的phaco,相变尖端温度保持在<28.0C。在关闭灌溉的情况下,相较于在灌溉条件下,温度更高,但仍保持在<45.0C。比较在开启灌溉时三种功率调制的温度,即在每个phaco功率和占空比设置中,连续phaco时phaco尖端的温度最高,其次是上升时间为1的超脉冲,然后是上升时间为2的超脉冲。关闭灌溉时,使用上升的超脉冲记录最高温度时间2,接着是连续的phaco,然后是具有上升时间1的超脉冲。与开启灌溉时的连续模式相比,超脉冲和波形调制减少了热量的产生。较低的占空比和较低的超声功率会在phaco尖端产生较少的热量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号