Effect of transmission losses on measured parameters in multi-ion coincidence momentum spectrometers

机译:传输损耗对多离子重合动量谱仪中测量参数的影响

获取原文
获取原文并翻译 | 示例

摘要

We address the effect of transmission losses on the measured kinematic parameters of molecular fragmentation processes in a time-of-flight based ion momentum spectrometer. We suggest means to correct for these losses for two-body fragmentation having isotropic distribution. As an example, we discuss the kinetic energy release spectrum observed in recoil ion momentum spectrometer of doubly ionized nitrogen molecules formed upon high energy electron impact. Further, we highlight the difficulties in calculating losses in many-body fragmentation and anisotropic emissions and show that such effects cannot be accounted without employing additional information about the process. In many cases, such information cannot be extracted from the observed data in the experiment and thus we have to rely on careful assumptions.
机译:我们在基于飞行时间的离子动量谱仪中研究了传输损耗对分子碎片化过程中测得的运动学参数的影响。对于具有各向同性分布的两体破碎,我们提出了纠正这些损失的方法。例如,我们讨论了在高能电子撞击后形成的双电离氮分子的反冲离子动量谱仪中观察到的动能释放谱。此外,我们强调了计算多体碎片和各向异性排放损失的困难,并表明,如果不使用有关该过程的其他信息,就无法考虑这种影响。在许多情况下,无法从实验中观察到的数据中提取此类信息,因此我们必须依靠仔细的假设。