A metastable atom bombardment (MAB) ionization source has been coupled to an existing thermal desorption aerosol mass spectrometer. The design allows real-time alternation between MAB and electron ionization (EI). A jet of metastable species produced in a DC discharge is directed at the ionization volume of the mass spectrometer, where Penning ionization is thought to be the dominant mechanism. Performance is characterized in experiments with oleic acid particles. By changing discharge gases between N2, Kr, and Ar, the excited state energy of the metastable species can be adjusted in the range 8.5-11.7 eV. For vaporization at 180 °C, all gases yield significantly less fragmentation than EI, which could improve results of factor analysis. Fragmentation increases with vaporization temperature, but generated fragments have higher average mass than those produced by EI. Analyte signal levels are 0.1% and 0.006% of equivalent analysis with EI when using Ar* and Kr*, respectively. These sensitivities are not practical for ambient studies, but are sufficient for source measurements, as demonstrated with direct measurements of biomass burning emissions. The measured Ar * flux of 3.6 × 1013 sr-1 s -1 is ~30 times lower than the best literature values for similar metastable beam sources, suggesting that sensitivity can be increased by source design improvements.
机译:亚稳态原子轰击(MAB)电离源已与现有的热脱附气溶胶质谱仪耦合。该设计允许在MAB和电子电离(EI)之间进行实时交替。在直流放电中产生的亚稳态物质的射流直接对准质谱仪的电离体积,在此认为Penning电离是主要机理。在油酸颗粒的实验中表征了性能。通过在N2,Kr和Ar之间改变放电气体,可以将亚稳态物质的激发态能量调节在8.5-11.7 eV的范围内。对于在180°C下汽化,所有气体产生的碎片均比EI少得多,这可以改善因子分析的结果。碎片随汽化温度的增加而增加,但生成的碎片的平均质量高于EI生成的碎片。当使用Ar *和Kr *时,分析物信号水平分别为EI等效分析的0.1%和0.006%。这些敏感性对于环境研究不切实际,但对于源测量已足够,如直接测量生物质燃烧排放所证明的那样。测量的Ar *通量为3.6×1013 sr-1 s -1,比同类亚稳光束源的最佳文献值低约30倍,这表明可以通过改进光源设计来提高灵敏度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号