Development of an imaging-planning program for screen/film and computed radiography mammography for breasts with short chest wall to nipple distance.

机译:为屏幕/胶片和胸部X线到乳头距离短的乳房的X线计算机摄影X线摄影成像计划的开发。

获取原文
获取原文并翻译 | 示例
       

摘要

OBJECTIVE: Imaging breasts with a short chest wall to nipple distance (CWND) using a traditional mammographic X-ray unit is a technical challenge for mammographers. The purpose of this study is the development of an imaging-planning program to assist in determination of imaging parameters of screen/film (SF) and computed radiography (CR) mammography for short CWND breasts. METHODS: A traditional mammographic X-ray unit (Mammomat 3000, Siemens, Munich, Germany) was employed. The imaging-planning program was developed by combining the compressed breast thickness correction, the equivalent polymethylmethacrylate thickness assessment for breasts and the tube loading (mAs) measurement. Both phantom exposures and a total of 597 exposures were used for examining the imaging-planning program. RESULTS: Results of the phantom study show that the tube loading rapidly decreased with the CWND when the automatic exposure control (AEC) detector was not fully covered by the phantom. For patient exposures with the AEC fully covered by breast tissue, the average fractional tube loadings, defined as the ratio of the predicted mAs using the imaging-planning program and mAs of the mammogram, were 1.10 and 1.07 for SF and CR mammograms, respectively. The predicted mAs values were comparable to the mAs values, as determined by the AEC. CONCLUSION: By applying the imaging-planning program in clinical practice, the experiential dependence of the mammographer for determination of the imaging parameters for short CWND breasts is minimised.
机译:目的:使用传统的乳腺X线摄影机对胸壁至乳头距离(CWND)短的乳房进行成像是乳腺X线摄影人员的一项技术挑战。这项研究的目的是开发一种成像计划程序,以帮助确定短CWND乳房的屏幕/胶片(SF)和计算机X线摄影(CR)乳腺摄影的成像参数。方法:采用传统的乳腺X射线照相设备(Mammomat 3000,西门子,慕尼黑,德国)。通过结合压缩的乳房厚度校正,等效的聚甲基丙烯酸甲酯乳房厚度评估和管负荷(mAs)测量,开发了成像计划程序。幻像曝光和总共597次曝光均用于检查成像计划。结果:体模研究的结果表明,当自动曝光控制(AEC)检测器没有完全被体模覆盖时,CWND可使管负载迅速降低。对于暴露在乳腺组织完全覆盖的AEC的患者暴露中,SF和CR乳腺X线照片的平均分数管负载分别为1.10和1.07,定义为使用成像计划程序预测的mAs与乳房X线照片的mAs之比。预测的mAs值与AEC确定的mAs值相当。结论:通过在临床实践中应用成像计划程序,乳腺X线摄影人员确定短CWND乳房成像参数的经验依赖性得以最小化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号