Patient doses from barium meal and barium enema examinations and potential for reduction through proper set-up of equipment.

机译:钡餐和钡灌肠检查的患者剂量以及通过适当设置设备而减少的可能性。

获取原文
获取原文并翻译 | 示例
           

摘要

Patient doses for barium meal and barium enema examinations, performed at two Greek hospitals, were measured using a dose-area product meter. The results were analysed to obtain the contributions of fluoroscopy and radiography to the dose as well as a number of other dose related parameters for each examination. The doses observed are within the range of values reported by other authors and comply with the dose reference levels (DRLs), proposed from relevant surveys in the UK and The Netherlands. However, comparison between the two hospitals revealed significant differences in the contributions to dose from the various parts of the examinations. To determine the reasons for these differences, measurements of dose related parameters were made using a Plexiglas phantom and standard clinical X-ray machine settings. Factors contributing to increased dose delivery were determined and recommendations have been made concerning ways in which doses might be reduced in each hospital, without degradation of the diagnostic quality of these examinations.
机译:在希腊两家医院进行的钡餐和钡灌肠检查的患者剂量使用剂量面积乘积计进行测量。分析结果以获得每次检查的透视和射线照相对剂量以及许多其他剂量相关参数的贡献。观察到的剂量在其他作者报告的值的范围内,并且符合英国和荷兰的相关调查所建议的剂量参考水平(DRL)。但是,两家医院之间的比较显示,各个检查部分对剂量的贡献存在显着差异。为了确定这些差异的原因,使用有机玻璃体模和标准的临床X射线机设置进行剂量相关参数的测量。确定了导致增加剂量输送的因素,并提出了关于在不降低这些检查的诊断质量的情况下,在每家医院中可以减少剂量的方式的建议。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号