Characteristics and performance evaluation of the capacitive Contact Flexible Microstrip Applicator operating at 70 MHz for external hyperthermia.

机译:用于外部热疗的70 MHz电容式接触式柔性微带敷料器的特性和性能评估。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

PURPOSE: To characterise and evaluate the capacitive Contact Flexible Microstrip Applicator operating at 70 MHz, CFMA-70. This applicator is introduced for the treatment of superficial tumours with extension in depth beyond the range of regular superficial applicators. METHODS: E-field measurements were performed in an elliptical phantom filled with a saline solution using an E-field vector probe. E-field distributions and SAR patterns are compared to those of the CFMA-434 and of 70 MHz waveguides. The applicator has been used for the treatment of 6 patients with breast cancer with a tumour depth exceeding 4 cm. RESULTS: The effective heating depth of the CFMA-70 is 50% larger than for the CFMA-434. Its effective field size is 26 x 20 cm (aperture 29 x 20 cm), larger than for an equally sized CFMA-434. In contrast to the CFMA-434 the SAR pattern of this applicator is insensitive to the bolus thickness. Comparison to 70 MHz waveguides shows that the E-field component normal to the applicator is 100% larger for the CFMA-70. During clinical applications acceptable temperatures were realised for individual sessions (also at depth), but in many cases treatment limiting hot spots occurred close to superficial bony structures near the applicator edge. Both surface irregularities and the normal field component may be responsible. CONCLUSIONS: The CFMA-70 has adequate penetration depth for the treatment of superficial tumours exceeding a depth of 4 cm. However, the relatively large normal E-field component may induce treatment-limiting hot spots at tissue interfaces in the direction normal to the applicator's surface.
机译:用途:表征和评估以70 MHz运行的电容式接触式柔性微带敷料器CFMA-70。介绍了这种涂药器,用于治疗浅表肿瘤,其深度扩展到常规浅表涂药器的范围之外。方法:使用电场矢量探针在装有盐溶液的椭圆形幻象中进行电场测量。将电场分布和SAR模式与CFMA-434和70 MHz波导的分布进行比较。该涂药器已用于治疗6名肿瘤深度超过4厘米的乳腺癌患者。结果:CFMA-70的有效加热深度比CFMA-434大50%。它的有效视场大小为26 x 20厘米(孔径为29 x 20厘米),大于同等大小的CFMA-434。与CFMA-434相比,此涂药器的SAR图案对弹丸厚度不敏感。与70 MHz波导的比较表明,对于CFMA-70,垂直于施加器的电场分量要大100%。在临床应用过程中,各个疗程(甚至在深处)均达到了可接受的温度,但在许多情况下,治疗限制热点出现在靠近施药器边缘的浅表骨结构附近。表面不规则和法向场分量都可能是造成这种情况的原因。结论:CFMA-70具有足够的穿透深度,可用于治疗深度超过4 cm的浅表肿瘤。但是,相对较大的法向电场分量可能会在组织界面的垂直于施药器表面的方向上引起限制治疗的热点。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号