Comparison of the clinical effectiveness of the 433 MHz Lucite cone applicator with that of a conventional waveguide applicator in applications of superficial hyperthermia.

机译:433 MHz萤石锥头涂药器与常规波导涂药器在浅层热疗应用中的临床效果比较。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: This report presents the final stage of our program to improve the quality of our superficial hyperthermia treatments. We have already demonstrated that the Lucite cone applicator (LCA), our technically improved water-filled, wave-guide applicator (WGA), is superior to the conventional WGA. The main objective of the present study was to investigate whether the technical improvements of a WGA were reflected in an improved clinical performance, e.g., a better temperature distribution. METHODS AND MATERIALS: Power and temperature analyses were performed retrospectively on 128 treatments of superficially located tumors (less than 4 cm depth). Twenty-three patients were treated alternately with a WGA setup and a LCA setup. RESULTS: The average power level per antenna in an array was 48 W and 62 W for the WGA and LCA respectively. The average invasively measured temperatures increased by 0.27 degrees C when the LCAs were used. The temperature difference between the center and the periphery of an antenna, averaged over the complete array of antennae, was 0.43 degrees C using WGAs and -0.05 degrees C using LCAs indicating a more uniform heating. The T90 of the invasively measured temperatures remained unchanged (WGA: 39.4 degrees C versus LCA: 39.5 degrees C). CONCLUSION: The LCA is now our standard applicator for superficial hyperthermia treatments as it is technically and clinically proven to be superior to the WGA.
机译:目的:本报告介绍了我们计划的最后阶段,以提高我们浅表热疗治疗的质量。我们已经证明,Lucite锥形涂抹器(LCA)是我们经过技术改进的注水波导涂抹器(WGA),优于传统的WGA。本研究的主要目的是研究WGA的技术改进是否反映在临床性能的改善(例如更好的温度分布)中。方法和材料:回顾性分析了浅表性肿瘤(深度小于4 cm)的128种治疗方法的功率和温度。 23例患者接受WGA和LCA交替治疗。结果:WGA和LCA阵列中每个天线的平均功率分别为48 W和62W。使用LCA时,平均侵入测量温度增加了0.27摄氏度。使用WGA,在整个天线阵列上平均的天线中心和外围之间的温差为0.43摄氏度,使用LCA为-0.05摄氏度,表明加热更为均匀。侵入测量温度的T90保持不变(WGA:39.4摄氏度,LCA:39.5摄氏度)。结论:LCA现在是我们用于浅层热疗的标准涂药器,因为它在技术和临床上均优于WGA。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号