Aim To assess the accuracy and reproducibility of biometry undertaken with the Aladdin (Topcon, Tokyo, Japan) in comparison with the current gold standard device, the IOLMaster 500 (Zeiss, Jena, Germany). Setting University Eye Clinic, Birmingham, UK and Refractive Surgery Centre, Kiel, Germany. Methods The right eye of 75 patients with cataracts and 22 healthy participants were assessed using the two devices. Measurements of axial length (AL), anterior chamber depth (ACD) and keratometry (K) were undertaken with the Aladdin and IOLMaster 500 in random order by an experienced practitioner. A second practitioner then obtained measurements for each participant using the Aladdin biometer in order to assess interobserver variability. Results No statistically significant differences ( p≥0.05) between the two biometers were found for average difference (AL)±95% CI=0.01±0.06 mm), ACD (0.00 ±0.11 mm) or mean K values (0.08±0.51 D). Furthermore, interobserver variability was very good for each parameter (weighted κ≥0.85). One patient's IOL powers could not be calculated with either biometer measurements, whereas a further three could not be analysed by the IOLMaster 500. The IOL power calculated from the valid measurements was not statistically significantly different between the biometers (p=0.842), with 91% of predictions within±0.25 D. Conclusions The Aladdin is a quick, easy-to-use biometer that produces valid and reproducible results that are comparable with those obtained with the IOLMaster 500.
机译:目的为了评估与当前的金标准设备IOLMaster 500(德国耶拿的蔡司)相比,使用Aladdin(日本东京的Topcon,日本东京)进行生物测定的准确性和可重复性。英国伯明翰的大学眼科诊所和德国基尔的屈光手术中心。方法使用这两种设备评估75例白内障患者和22名健康参与者的右眼。有经验的医生以随机顺序使用Aladdin和IOLMaster 500进行轴向长度(AL),前房深度(ACD)和角膜曲率(K)的测量。然后,第二位从业者使用阿拉丁生物测定仪获得了每个参与者的测量结果,以评估观察者之间的差异。结果两种生物测定仪之间的平均差异(AL)±95%CI = 0.01±0.06 mm),ACD(0.00±0.11 mm)或平均K值(0.08±0.51 D)没有发现统计学上的显着差异(p≥0.05) 。此外,每个参数的观察者间差异非常好(加权κ≥0.85)。两种生物测量仪均无法计算一名患者的IOL功效,而IOLMaster 500则无法分析另外三名患者。根据有效测量结果计算出的IOL功效在各生物计之间无统计学差异(p = 0.842),其中91预测值的百分比在±0.25 D之内。结论Aladdin是一种快速,易于使用的生物测定仪,可产生有效且可重复的结果,可与IOLMaster 500获得的结果相媲美。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号