A Detector for the Pulsed Nuclear Quadrupole Resonance Spectrometer

机译:脉冲核四极共振光谱仪的检测器

获取原文
获取原文并翻译 | 示例
           

摘要

An increase in the sensitivity of a pulse-type detector for the nuclear quadrupole resonance spectrometer has been attained by optimizing the delay time for switching on the receiver after termination of an excitation pulse and by using an original circuit of the synchronous detector with a high dynamic range of 100 dB. The sensitivity of the nuclear quadrupole resonance spectrometer improved by using the developed detector is about twice as great as the spectrometer sensitivity before its upgrading. The spectrometer is capable of detecting 0.5 g of sodium nitrite.
机译:通过优化激励脉冲终止后接通接收器的延迟时间并使用具有高动态特性的同步检测器的原始电路,可以提高用于核四极共振光谱仪的脉冲型检测器的灵敏度。范围为100 dB。通过使用开发的检测器提高的核四极共振光谱仪的灵敏度大约是升级前光谱仪灵敏度的两倍。该光谱仪能够检测0.5 g亚硝酸钠。