A radiation foil bolometer intended for measuring radiation-energy fluxes and plasma flows in high-power pulsed plasma X-ray sources is studied. In the bolometer, the radiation or ion-flow energy is absorbed by a 13-μm-thick niobium foil preliminarily heated to T≈1900 K by a quasi-constant current lasting 2 s. Radiation from the foil was fed through an 8-m-long light guide to a semiconductor photodetector. The sensitivity of the instrument obtained from the results of calibrations is 0.56 J/Vcm~2. The time resolution of this technique (~2.5 μs) is determined by the foil heating time. In experiments with mega-ampere Z-pinches, the time resolution attained allows detection of the radiation energy independently of the energy of expanding plasma flows. In this case, the photoelectric effect from X-rays and the conductivity of expanding plasma do not affect the operation of the radiation bolometer. The working capacity of the bolometer was demonstrated at the Angara-5-1 facility in experiments with high-current Z-pinches with a radiation output energy of 50 - 100 kJ/pulse. The measurements performed have shown that the energy flux density of plasma expanding in the direction perpendicular to the pinch axis is at most 5% of the energy flux density of soft X-rays.