Haemoglobin A1c: evaluation of a new HbA1c point-of-care analyser Bio-Rad in2it in comparison with the DCA 2000 and central laboratory analysers.

机译:血红蛋白A1c:与DCA 2000和中央实验室分析仪相比,对新型HbA1c即时分析仪Bio-Rad in2it进行评估。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Point-of-care-testing (POCT) of haemoglobin Alc (HbA1c) is popular due to its fast turnaround of results in the outpatient setting. The aim of this project was to evaluate the performance of a new HbA1c POCT analyser, the Bio-Rad in2it, and compare it with the Siemens DCA 2000, Bio-Rad Variant II and Roche Tina-quant HbA1c Gen 2 assay on the cobas c501. METHODS: Imprecision of the four methods were compared by computing total imprecision from within-run and between-run data. A total of 80 samples were also compared and analysed by Deming regression and Altman-Bland difference test. RESULTS: Study of total imprecision of the in2it at HBA1c levels of 6.0% and 10.4% produced a coefficient of variation (%CV) of 3.8% and 3.7%, respectively. These results were more favourable as compared with the DCA 2000 but did not match the low imprecision of the central laboratory methods, the Bio-Rad Variant II and the Roche cobas c501. Comparison between the in2it and the central laboratory analysers, Bio-Rad variant II and cobas c501, revealed positive bias of 12% and 10%, respectively, supported by corresponding Deming regression equation slopes of +1.18 and +1.14. Comparison between the DCA 2000 and the central laboratory analysers revealed a bias that became increasingly positive with rising HbA1c concentrations with Deming regression analysis also revealing proportional and constant differences. CONCLUSIONS: The in2it is a suitable POCT analyser for HbA1c but its less than ideal precision performance and differences with the central laboratory analysers must be communicated to and noted by the users.
机译:背景:血红蛋白Alc(HbA1c)的即时检验(POCT)受欢迎,因为它在门诊环境中的结果快速周转。该项目的目的是评估新型HbA1c POCT分析仪Bio-Rad in2it的性能,并将其与cobas c501上的Siemens DCA 2000,Bio-Rad Variant II和Roche Tina-quant HbA1c Gen 2分析进行比较。方法:通过计算运行中和运行间数据的总不精确度,比较了四种方法的不精确度。还通过Deming回归和Altman-Bland差异检验对总共80个样本进行了比较和分析。结果:在HBA1c水平为6.0%和10.4%的情况下,对in2it的总不精确度进行了研究,得出的变异系数(%CV)分别为3.8%和3.7%。与DCA 2000相比,这些结果更为令人满意,但与中央实验室方法,Bio-Rad Variant II和Roche cobas c501的低精确度不匹配。在in2it和中央实验室分析仪Bio-Rad II和cobas c501之间进行的比较显示,正偏倚分别为12%和10%,相应的Deming回归方程斜率分别为+1.18和+1.14。 DCA 2000和中央实验室分析仪之间的比较显示,随着HbA1c浓度的升高,偏倚正变得越来越正。戴明回归分析也显示出比例和常数差异。结论:in2it是适用于HbA1c的POCT分析仪,但其精度性能不理想,并且必须与用户沟通并注意与中央实验室分析仪的差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号