首页> 外文期刊>Annals of clinical biochemistry. >Effect of non-glucose sugars and haematocrit on glucose measurements with Roche Accu-Chek Performa glucose strips.
【24h】

Effect of non-glucose sugars and haematocrit on glucose measurements with Roche Accu-Chek Performa glucose strips.

机译:非葡萄糖糖和血细胞比容对使用Roche Accu-Chek Performa葡萄糖条进行葡萄糖测量的影响。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Point-of-care blood glucose meters based on glucose dehy-drogenase (GDH) and pyrroloquinoline quinone (PQQ)-coupled reaction are subject to interference by non-glucose sugars and icodextrin metabolites, chiefly maltose.Reports of treatment of spurious hyperglycaemia, arising from the interference, with insulin leading to life-threatening/fatal hypoglycaemia caused the Food and Drug Administration and others to issue safety notices to alert health professionals of the dangers of using GDH-PQQ-based glucose monitoring systems.
机译:基于葡萄糖脱氢酶(GDH)和吡咯并喹啉醌(PQQ)偶联反应的即时血糖仪会受到非葡萄糖糖和艾考糊精代谢产物(主要是麦芽糖)的干扰。由于存在干扰,胰岛素导致生命危险/致命性低血糖,导致美国食品药品监督管理局(FDA)和其他机构发布安全公告,以警告健康专业人员使用基于GDH-PQQ的葡萄糖监测系统的危险。