Field-testing and validation in a primary care setting of a point-of-care test for C-reactive protein.

机译:在初级保健机构中对C反应蛋白进行即时检验的现场测试和验证。

获取原文
获取原文并翻译 | 示例

摘要

BACKGROUND: Use of point-of-care testing (POCT) in primary care has increased. There is a need for high-quality field evaluation of POCT before deployment can be considered. METHOD: A POCT system for C-reactive protein was evaluated in a routine general practice setting. The standard laboratory method was a dry slide method based in a routine hospital laboratory. RESULTS: Scatterplots for both venous and capillary blood POCT system results versus the standard laboratory assay produced correlation coefficients of greater than 0.96. Bland-Altman plots indicated that 95% of venous and capillary POCT results fell within +/-10 mg/L of the mean value with no clinically significant difference from laboratory results. CONCLUSIONS: The POCT system performed reliably in a routine general practice setting with satisfactory performance against an accepted laboratory method.
机译:背景:在初级保健中使用即时医疗点检测(POCT)的情况有所增加。在考虑部署之前,需要对POCT进行高质量的现场评估。方法:在常规的常规操作环境中评估了C反应蛋白的POCT系统。标准实验室方法是基于常规医院实验室的干玻片方法。结果:散点图的静脉和毛细血管POCT系统结果与标准实验室分析相比,相关系数大于0.96。 Bland-Altman图表明,静脉和毛细血管POCT结果的95%落在平均值的+/- 10 mg / L之内,与实验室结果没有临床上的显着差异。结论:POCT系统在常规的常规操作环境中可靠地运行,并且相对于公认的实验室方法具有令人满意的性能。