We are reporting a case of severe haemolytic anemia with cold agglutinins which combines several spurious determinations. It shows the usefulness of the new erythrocytic parameters of the XE 5000 Sysmex, specially: red blood cells with optical count (RBC-O), GR-He (intra-erythocytic hemoglobin) and R-MFV (most frequent volume). Optical red blood cells act as a substitute for red cells count instead of impedance red cells and R-MFV as a substitute for MCV (mean cell volume). The hematocrit (HCT) is corrected thanks to the following formula: HCT=(RBC-O X R- MFV)/1000. Free plasmatic hemoglobin is included in the measure of hemoglobin by the analyzer but is not available for tissue oxygenation. So, hemoglobin (HGB) has to be corrected by the means of GR-He thanks to the following formula: HGB=(GR He x RBC-O)/10.
机译:我们正在报告一例严重的溶血性贫血伴冷凝集素,结合了一些虚假的判断。它显示了XE 5000 Sysmex新的红细胞参数的实用性,特别是:具有光学计数的红细胞(RBC-O),GR-He(红细胞内血红蛋白)和R-MFV(最频繁的体积)。光学红细胞代替红细胞计数代替阻抗红细胞,而R-MFV代替MCV(平均细胞体积)。通过以下公式校正血细胞比容(HCT):HCT =(RBC-O X R-MFV)/ 1000。分析仪测量血红蛋白时会包含游离血浆血红蛋白,但组织氧合不可用。因此,由于以下公式,必须通过GR-He校正血红蛋白(HGB):HGB =(GR He x RBC-O)/ 10。