首页> 外文期刊>Arthroscopy: the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association >Hybrid femoral fixation of soft-tissue grafts in anterior cruciate ligament reconstruction using the EndoButton CL and bioabsorbable interference screws: a biomechanical study.
【24h】

Hybrid femoral fixation of soft-tissue grafts in anterior cruciate ligament reconstruction using the EndoButton CL and bioabsorbable interference screws: a biomechanical study.

机译:使用EndoButton CL和可生物吸收的干扰螺钉在前交叉韧带重建中混合软组织移植物的股骨固定:一项生物力学研究。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: The purpose of this study was to evaluate the effect of hybrid femoral fixation with bioabsorbable interference screws (BioRCI; Smith & Nephew Endoscopy, Andover, MA) and EndoButton CL (Smith & Nephew Endoscopy) fixation. METHODS: Biomechanical testing of 3 different fixation techniques was performed by use of porcine hind-limb distal femurs and mature bovine extremity common extensor tendons. Two independent testing sessions were examined. The first testing session (group A) compared femoral fixation via the EndoButton CL device (n = 6) with femoral fixation via the EndoButton CL device with the addition of a BioRCI screw (n = 6). The second testing session (group B) compared femoral fixation via BioRCI screws alone (n = 6) with femoral fixation via the EndoButton CL device with the addition of a BioRCI screw (n = 6). The femur-graft complex was cyclically loaded between 50 and 250 N at 1 Hz for 1,000 cycles. After cycling, the amount of graft slippage was determined by measuring the change in grip-to-grip distance. The complex was then loaded to failure at 1 mm/s, and the ultimate tensile strength, stiffness, and mode of failure were determined. RESULTS: In group A the addition of an interference screw to the EndoButton CL fixation increased the ultimate tensile strength (1,364.7 +/- 102.4 N for EndoButton CL alone v 1,449.3 +/- 94.4 N for combined technique, P = .035) and stiffness (195.5 +/- 12.1 N/mm for EndoButton CL alone v 307.3 +/- 54.9 N/mm for combined technique, P = .004) and decreased the amount of graft slippage (2.6 +/- 0.5 mm for EndoButton CL alone v 2.0 +/- 0.3 mm for combined technique, P = .017). In group B the addition of the EndoButton CL device to interference screw fixation significantly increased the ultimate tensile strength (643.5 +/- 148.4 N for BioRCI screws alone v 1,290.3 +/- 254.4 N for combined technique, P = .004) but had no effect on stiffness (315.7 +/- 38.9 N/mm for BioRCI screws alone v 341.5 +/- 64.0 N/mm for combined technique, P = .267) or graft slippage (2.7 +/- 1.0 mm for BioRCI screws alone v 2.0 +/- 0.6 mm for combined technique, P = .087). CONCLUSIONS: Our study shows that hybrid femoral fixation of double-looped gracilis-semitendinosus grafts via the EndoButton CL device and a bioabsorbable interference screw is stronger than interference or EndoButton CL fixation alone with respect to ultimate tensile strength, stiffness, and slippage. The addition of an interference screw to suspensory fixation via the EndoButton CL device increased the ultimate tensile strength from 1,360 N to 1,450 N, improved reconstruction stiffness from 200 N/mm to 300 N/mm, and decreased the amount of graft slippage resulting from cyclic loading from 2.6 mm to 2.0 mm. CLINICAL RELEVANCE: The hybrid fixation of the EndoButton CL device and an interference screw is a stronger and stiffer construct than either device alone and allows for aperture fixation, which may translate into better clinical results.
机译:目的:本研究的目的是评估采用生物可吸收干扰螺钉(BioRCI; Smith&Nephew内窥镜检查,安德弗,马萨诸塞州)和EndoButton CL(Smith&Nephew内窥镜检查)进行混合股骨固定的效果。方法:通过使用猪后肢远端股骨和成熟的牛四肢普通伸肌腱进行三种固定技术的生物力学测试。检查了两个独立的测试会话。第一次测试(A组)比较了通过EndoButton CL设备进行股骨固定(n = 6)与通过EndoButton CL设备进行股骨固定以及BioRCI螺钉(n = 6)的比较。第二个测试阶段(B组)比较了仅通过BioRCI螺钉进行股骨固定(n = 6)与通过EndoButton CL装置通过增加BioRCI螺钉进行股骨固定(n = 6)。将股骨移植物复合物以1 Hz的频率在50和250 N之间循环加载1,000个循环。循环后,通过测量抓地到抓地距离的变化来确定移植物的滑动量。然后将复合体以1 mm / s的速度加载至破坏,并确定极限拉伸强度,刚度和破坏模式。结果:在A组中,在EndoButton CL固定件上增加了干涉螺钉,从而提高了极限拉伸强度(单独使用EndoButton CL时为1,364.7 +/- 102.4 N,联合技术为1,449.3 +/- 94.4 N,P = .035)和刚度(单独使用EndoButton CL时为195.5 +/- 12.1 N / mm v对于组合技术而言为307.3 +/- 54.9 N / mm,P = .004)并减少了移植物打滑量(单独使用EndoButton CL时为2.6 +/- 0.5 mm v组合技术为2.0 +/- 0.3毫米,P = .017)。在B组中,在干涉螺钉固定中添加EndoButton CL装置显着提高了极限抗拉强度(仅BioRCI螺钉为643.5 +/- 148.4 N,联合技术为1,290.3 +/- 254.4 N,P = .004),但没有对刚度的影响(对于单独的BioRCI螺钉而言为315.7 +/- 38.9 N / mm对于组合技术为v 341.5 +/- 64.0 N / mm,P = .267)或移植物的滑动(对于单独的BioRCI螺钉而言为2.7 +/- 1.0 mm对v 2.0组合技术时为+/- 0.6毫米,P = .087)。结论:我们的研究表明,就最终抗张强度,刚度和滑移而言,通过EndoButton CL装置和可生物吸收的干扰螺钉对双环gra肌-半end肌的混合股骨固定比单独使用干涉或EndoButton CL固定更为牢固。通过EndoButton CL装置在悬吊固定系统中增加了干涉螺钉,从而将极限抗拉强度从1,360 N提高至1,450 N,将重建刚度从200 N / mm提高至300 N / mm,并减少了因循环引起的移植物滑移量加载范围从2.6毫米到2.0毫米。临床相关性:EndoButton CL设备和干涉螺钉的混合固定结构比任何一个单独的设备都更坚固,更坚固,并且可以进行光阑固定,这可以转化为更好的临床效果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号