Locking plates have increased torsional stiffness compared to standard plates in a segmental defect model of clavicle fracture.

机译:与锁骨骨折节段性缺损模型中的标准板相比,锁定板具有更高的扭转刚度。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

OBJECTIVE: To evaluate the effect of locked plate technology to resist torsion in a clavicle fracture model of segmental bone loss. METHODS: Forty-four synthetic clavicles were repaired with either 3.5 mm locked compression plate (LCP) or 3.5 mm low-contact dynamic compression plate (LCDCP). They were divided into two groups of 22 specimens. Each group was tested to evaluate torsional stiffness, load at failure, deflection at failure, and unconstrained plate motion. RESULTS: LCP group showed significantly greater stiffness in torsion compared to the LCDCP group (p < 0.001). Average difference was 20.9%. Load at failure was not significantly different (p < 0.07). Deflection at failure was significantly less for the LCP group (p < 0.03). Unconstrained motion or plate 'looseness' was significantly less for the LCP group (p < 0.017). CONCLUSIONS: In a simulated model of segmental clavicle fracture, a LCP provided more stiffness and less deflection than a low-contact dynamic compression plate.
机译:目的:评价锁骨板技术在节段性骨丢失的锁骨骨折模型中抗扭转的作用。方法:用3.5 mm锁定加压板(LCP)或3.5 mm低接触动态加压板(LCDCP)修复了44个合成锁骨。他们分为22个标本的两组。每组均经过测试以评估扭转刚度,破坏时的载荷,破坏时的挠度以及不受约束的板运动。结果:与LCDCP组相比,LCP组显示出明显的扭转刚度(p <0.001)。平均差异为20.9%。失效负载无显着差异(p <0.07)。 LCP组在衰竭时的挠曲明显更少(p <0.03)。 LCP组的运动不受约束或板块“松动”明显更少(p <0.017)。结论:在节段性锁骨骨折的模拟模型中,LCP比低接触动态压迫板具有更大的刚度和更小的挠度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号