Validation of an iPad test of letter contrast sensitivity

机译:iPad的验证测试对比的信灵敏度

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: An iPad-based letter contrast sensitivity test was developed (ridgevue.com) consisting of two letters on each page of an iBook. The contrast decreases from 80% (logCS = 0.1) to 0.5% (logCS = 2.3) by 0.1 log units per page. The test was compared to the Pelli-Robson Test and the Freiburg Acuity and Contrast Test. METHODS: Twenty normally sighted subjects and 20 low-vision subjects were tested monocularly at 1 m using each test wearing their habitual correction. After a 5-minute break, subjects were retested with each test in reverse order. Two different letter charts were used for both the Pelli-Robson and iPad tests, and the order of testing was varied systematically. For the Freiburg test, the target was a variable contrast Landolt C presented at eight possible orientations and used a 30-trial Best PEST procedure. Repeatability and agreement were assessed by determining the 95% limits of agreement (LoA) ±1.96 SD of the differences between administrations or tests. RESULTS: All three tests showed good repeatability in terms of the 95% LoA: iPad = ±0.19, Pelli-Robson = ±0.19, and Freiburg = ±0.15. The iPad test showed good agreement with the Freiburg test with similar mean (±SD) logCS (iPad = 1.98 ± 0.11, Freiburg = 1.96 ± 0.06) and with narrow 95% LoA (±0.24), but the Pelli-Robson test gave significantly lower values (1.65 ± 0.04). Low-vision subjects had slightly poorer repeatability (iPad = ±0.24, Pelli-Robson = ±0.23, Freiburg = ±0.21). Agreement between the iPad and Freiburg tests was good (iPad = 1.45 ± 0.40, Freiburg = 1.54 ± 0.37), but the Pelli-Robson test gave significantly lower values (1.30 ± 0.30). CONCLUSIONS: The iPad test showed similar repeatability and may be a rapid and convenient alternative to some existing measures. The Pelli-Robson test gave lower values than the other tests.
机译:目的:对比敏感度iPad-based信测试开发(ridgevue.com)组成的两个字母在每个页面的iBook。对比降低(logCS = 0.1)的80%到0.5%(logCS = 2.3) 0.1日志单位每个页面。Pelli-Robson测试相比,弗莱堡敏锐度和对比试验。二十通常有视力的受试者和20低视力患者单眼1进行测试米使用每个测试穿着习惯修正。重新测试,每个测试在相反的顺序。不同字母图表的使用Pelli-Robson和iPad测试的顺序系统测试是不同的。弗莱堡测试,目标是一个变量的对比Landolt C提出了八点方向,用30-trial最好的害虫过程。评估确定95%的限制协议(贷款)±1.96 SD的差异政府或测试。三个测试方面表现出良好的可重复性和弗莱堡=±0.15。协议与弗莱堡测试相似意思是(±SD) logCS (iPad = 1.98±0.11,弗莱堡=1.96±0.06)和狭窄95%贷款(±0.24),但是Pelli-Robson测试给了显著降低值(1.65±0.04)。略穷可重复性(iPad =±0.24,iPad和弗莱堡测试之间的协议好(iPad = 1.45±0.40,弗莱堡= 1.54±0.37),但Pelli-Robson测试了显著降低值(1.30±0.30)。结论:iPad测试显示相似重复性和可能是一个快速和方便替代现有的一些措施。Pelli-Robson测试值低于其他测试。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号