Study of the IsoAid ADVANTAGE (125)I brachytherapy source dosimetric parameters using Monte Carlo simulation.

机译:使用Monte Carlo模拟研究IsoAid ADVANTAGE(125)I近距离放射源剂量参数。

获取原文
获取原文并翻译 | 示例
           

摘要

Monte Carlo calculations have been performed using the MCNP4C code for an iodine seed design. As the ADVANTAGE I-125, Model IAI-125 source is commercially available for interstitial brachytherapy treatment, dosimetric characteristics (dose rate constant, radial dose function and anisotropy function) of this source were theoretically determined following the updated AAPM task group 43 (TG-43U1) recommendations. The dose distribution around the seed was calculated with Monte Carlo simulation in liquid water. The Monte Carlo calculated dose rate constant of this source in water was found to be 0.986cGyh(-1)U(-1), with an approximate uncertainty of 0.4%. The obtained result has been compared with the previous study. Comparison of the calculated dose rate constant with the value presented by Meigooni et al. shows a very good agreement. Also the anisotropy function and the radial dose function for this source are graphically compared.
机译:蒙特卡洛计算已使用MCNP4C代码进行碘种子设计。由于IAI-125型ADVANTAGE I-125来源可用于间质近距离放射治疗,因此根据更新的AAPM任务组43(TG-G),从理论上确定了该来源的剂量特性(剂量率常数,径向剂量函数和各向异性函数) 43U1)建议。通过蒙特卡洛模拟在液态水中计算种子周围的剂量分布。蒙特卡洛计算得出的该源在水中的剂量率常数为0.986cGyh(-1)U(-1),不确定度约为0.4%。获得的结果已与以前的研究进行了比较。计算的剂量率常数与Meigooni等人提供的值的比较。显示出很好的协议。还以图形方式比较了该源的各向异性函数和径向剂量函数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号