Performance characteristics of a gated fiber-optic-coupled dosimeter in high-energy pulsed photon radiation dosimetry.

机译:门控光纤耦合剂量计在高能脉冲光子辐射剂量测定中的性能特征。

获取原文
获取原文并翻译 | 示例

摘要

Fiber-optic-coupled dosimeters (FOCDs) are a new class of in vivo dosimetry systems that are finding increased clinical applications. Utility of FOCDs has been limited in dosimetric applications due Cerenkov-ray signal contamination. The current study reports on the characterization of a novel FOCD, with a gated detection system for the discrimination and effective elimination of the direct contribution of Cerenkov radiation, for use in the radiotherapeutic realm. System reproducibility, linearity and output dependence on dose rate, energy, field size, and temperature response were characterized for 6, 10, and 15MV photon energies. The system exhibited a linear response to absorbed dose ranging from 1 to 2400cGy and showed little dependence to dose rate variations. Overall system reproducibility was 0.52% with no field-geometry and temperature dependence.
机译:光纤耦合剂量计(FOCD)是一类新型的体内剂量测定系统,正在发现越来越多的临床应用。由于切伦科夫射线信号污染,FOCD的用途在剂量学应用中受到限制。当前的研究报道了一种新型FOCD的特性,该系统具有用于检测和有效消除切伦科夫辐射直接贡献的门控检测系统,用于放射治疗领域。表征了6、10和15MV光子能量的系统重现性,线性和输出对剂量率,能量,场大小和温度响应的依赖性。该系统对吸收剂量范围为1至2400cGy表现出线性响应,对剂量率变化的依赖性很小。整个系统的重现性为0.52%,没有场几何和温度依赖性。