Initial forces and moments delivered by removable thermoplastic appliances during rotation of an upper central incisor.

机译:在上部中央门牙旋转期间,可移动的热塑性塑料装置传递的初始力和力矩。

获取原文
获取原文并翻译 | 示例

摘要

OBJECTIVE: To determine the forces and moments delivered to a maxillary central incisor during rotation by three different thermoplastic appliances with identical thickness. MATERIALS AND METHODS: Five identical appliances were manufactured from each of three materials (Ideal Clear 1.0 mm, Erkodur 1.0 mm, Biolon 1.0 mm). An upper central incisor fixed in a measuring device was rotated around its central axis in 0.5-degree steps to +/-2.5 degrees, +/-5 degrees, and +/-7.5 degrees (equivalent to an activation of +/-0.17 mm, +/-0.34 mm, and +/-0.51 mm of the incisor edge) in the clockwise and anticlockwise directions with the respective appliance fixed in place. For statistical analysis, the moments Tz (rotation) and forces Fz (intrusion) were tested. Means and standard deviations for Tz and median and 25% and 75% quantiles for Fz were calculated. An analysis of variance (ANOVA) was performed. RESULTS: The minimal moment was determined at a rotation of -0.17 mm (-7.3 Nmm, +/-0.8), and the maximal moment at a deflection of -0.51 mm (-71.8 Nmm. +/-2.5) was recorded. The minimal value for Fz was measured at an activation of -0.17 mm (0.0 N), and the highest intrusive forces were evaluated for a rotation of -0.51 mm (-5.8 N). The particular material sometimes had a significant (P < .05) influence on the forces delivered and the moments. CONCLUSIONS: During rotation with aligners, an intrusive force can also be observed. The direction of rotation, and the materials used all exert an influence on the force delivery properties of the appliance.
机译:目的:确定三种厚度相同的热塑性塑料在旋转过程中传递到上颌中切牙的力和力矩。材料与方法:用三种材料(理想的透明材料1.0毫米,Erkodur厚度1.0毫米,Biolon厚度1.0毫米)中的每一种制造了五个相同的器具。固定在测量设备中的上中切牙围绕其中心轴以0.5度的步长旋转到+/- 2.5度,+ /-5度和+/- 7.5度(相当于激活+/- 0.17毫米分别在顺时针和逆时针方向上将其分别切入(门牙边缘的+/- 0.34 mm和+/- 0.51 mm)。为了进行统计分析,测试了力矩Tz(旋转)和力Fz(侵入)。计算了Tz和中位数的平均值和标准偏差,以及Fz的25%和75%的分位数。进行了方差分析(ANOVA)。结果:最小力矩是在-0.17毫米(-7.3 Nmm,+ /-0.8)的旋转下确定的,最大力矩是在-0.51毫米(-71.8 Nmm。+/- 2.5)的偏转下记录的。 Fz的最小值是在-0.17 mm(0.0 N)的激活下测量的,而最高的插入力是在-0.51 mm(-5.8 N)的旋转下评估的。特定的材料有时会对传递的力和力矩产生重大(P <.05)影响。结论:在使用对准器旋转期间,还可以观察到侵入力。旋转方向以及所使用的材料都会对设备的力传递特性产生影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号