Analysis of automated external defibrillator device failures reported to the food and drug administration

机译:向食品药品监督管理局报告的自动体外除颤器设备故障分析

获取原文
获取原文并翻译 | 示例
           

摘要

Study objective: Automated external defibrillators are essential for treatment of cardiac arrest by lay rescuers and must determine when to shock and if they are functioning correctly. We seek to characterize automated external defibrillator failures reported to the Food and Drug Administration (FDA) and whether battery failures are properly detected by automated external defibrillators. Methods: FDA adverse event reports are catalogued in the Manufacturer and User Device Experience (MAUDE) database. We developed and internally validated an instrument for analyzing MAUDE data, reviewing all reports in which a fatality occurred. Two trained reviewers independently analyzed each report, and a third resolved discrepancies or passed them to a committee for resolution. Results: One thousand two hundred eighty-four adverse events were reported between June 1993 and October 2008, of which 1,150 were failed defibrillation attempts. Thirty-seven automated external defibrillators never powered on, 252 failed to complete rhythm analysis, and 524 failed to deliver a recommended shock. In 149 cases, the operator disagreed with the device's rhythm analysis. In 54 cases, the defibrillator stated the batteries were low and in 110 other instances powered off unexpectedly. Interrater agreement between reviewers 1 and 2 ranged by question from 69.0% to 98.6% and for most likely cause was 55.9%. Agreement was obtained for 93.7% to 99.6% of questions by the third reviewer. Remaining discrepancies were resolved by the arbitration committee. Conclusion: MAUDE information is often incomplete and frequently no corroborating data are available. Some conditions not detected by automated external defibrillators during self-test cause units to power off unexpectedly, causing defibrillation delays. Backup units frequently provide shocks to patients.
机译:研究目的:自动化体外除颤器对于非专业抢救者治疗心脏骤停是必不可少的,并且必须确定何时电击以及其功能是否正常。我们试图确定向食品和药物管理局(FDA)报告的自动外部除颤器故障的特征,以及自动外部除颤器是否能够正确检测到电池故障。方法:FDA不良事件报告在制造商和用户设备体验(MAUDE)数据库中分类。我们开发并内部验证了一种用于分析MAUDE数据,审查所有发生死亡的报告的工具。两名训练有素的审稿人独立分析了每个报告,第三项解决了差异或将其传递给委员会以解决。结果:1993年6月至2008年10月之间报告了124例不良事件,其中1,150例除颤失败。三十七台自动体外除颤器从未通电,252台未能完成心律分析,524台未能提供建议的电击。在149种情况下,操作员不同意设备的节奏分析。在54种情况下,除颤器显示电池电量低,在其他110种情况下意外断电。审稿人1和2之间的评分者间一致性从69.0%到98.6%不等,最可能的原因是55.9%。第三位审稿人对问题的93.7%至99.6%达成了共识。其余差异由仲裁委员会解决。结论:MAUDE信息通常不完整,并且经常没有确证的数据。自动外部除颤器在自检期间未检测到某些情况,导致设备意外关机,从而导致除颤延迟。备用单位经常给患者带来震惊。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号