A 59-year-old male who had undergone aortic and mitral valve replacement with Starr- Edwards ball valves 27 years ago was admitted to our hospital for hemolytic anemia and heart failure. Echocardiography revealed prosthetic valve failure with a high-pressure gradient and small effective orifice area. The Starr-Edwards ball valves were successfully replaced with bileaflet mechanical valves. The explanted valves revealed no structural abnormalities. ?2012 The Editorial Committee of Annals of Thoracic and Cardiovascular Surgery. All rights reserved.
机译:一位59岁的男性27年前接受了Starr-Edwards球阀的主动脉和二尖瓣置换手术,因溶血性贫血和心力衰竭入院。超声心动图显示人工瓣膜衰竭,高压梯度小,有效孔面积小。 Starr-Edwards球阀已被双叶式机械阀成功替代。外植瓣膜未发现结构异常。 2012年《胸心血管外科杂志》编辑委员会。版权所有。