The aim of this study was to evaluate the pupil size of normal subjects at different illumination levels with a novel pupillometer.The pupil size of healthy study participants was measured with an infrared-video PupilX pupillometer (MEye Tech GmbH, Alsdorf, Germany) at five different illumination levels (0, 0.5,4,32, and 250 lux). Measurements were performed by the same investigator. Ninety images were executed during a measurement period of 3 seconds. The absolute linear camera resolution was approximately 20 pixels per mm. This cross-sectional study analysed 490 eyes of 245 subjects (mean age: 51.9 ± 18.3 years, range: 6-87 years). On average, pupil diameter decreased with increasing light intensities for both eyes, with a mean pupil diameter of 5.39 ± 1.04 mm at 0 lux, 5.20 ± 1.00 mm at 0.5 lux, 4.70 ± 0.97 mm at 4 lux, 3.74 ± 0.78 mm at 32 lux, and 2.84 ± 0.50 mm at 250 lux illumination. Furthermore, it was found that anisocoria increased by 0.03 mm per life decade for all illumination levels (R~2 = 0.43). Anisocoria was higher under scotopic and mesopic conditions. This study provides additional information to the current knowledge concerning age- and light-related pupil size and anisocoria as a baseline for future patient studies.