首页> 外文期刊>Analytical Biochemistry: An International Journal of Analytical and Preparative Methods >Interference of some aqueous two-phase system phase-forming components in protein determination by the Bradford method
【24h】

Interference of some aqueous two-phase system phase-forming components in protein determination by the Bradford method

机译:布拉德福德法测定蛋白质时某些水两相系统相形成成分的干扰

获取原文
获取原文并翻译 | 示例
           

摘要

The interference of some specific aqueous two-phase system (ATPS) phase-forming components in bovine serum albumin (BSA) determination by the Bradford method was investigated. For this purpose, calibration curves were obtained for BSA in the presence of different concentrations of salts and polymers. A total of 19 salts [Na _2SO _4, (NH _4) _2SO _4, MgSO _4, LiSO _4, Na _2HPO _4, sodium phosphate buffer (pH 7.0), NaH _2PO _4, K _2HPO _4, potassium phosphate buffer (pH 7.0), KH _2PO _4, C _6H _8O _7, Na _3C _6H _5O _7, KCHO _2, NaCHO _2, NaCO _3, NaHCO _3, C _2H _4O _2, sodium acetate buffer (pH 4.5), and NaC _2H _3O _2] and 7 polymers [PEG 4000, PEG 8000, PEG 20000, UCON 3900, Ficoll 70000, PES 100000, and PVP 40000] were tested, and each calibration curve was compared with the one obtained for BSA in water. Some concentrations of salts and polymers had considerable effect in the BSA calibration curve. Carbonate salts were responsible for the highest salt interference, whereas citric and acetic acids did not produce interference even in the maximum concentration level tested (5 wt%). Among the polymers, UCON gave the highest interference, whereas Ficoll did not produce interference when used in concentrations up to 10 wt%. It was concluded that a convenient dilution of the samples prior to the protein quantification is needed to ensure no significant interference from ATPS phase-forming constituents.
机译:研究了使用Bradford方法测定牛血清白蛋白(BSA)中某些特定的水性两相系统(ATPS)相形成成分的干扰。为此,在存在不同浓度的盐和聚合物的情况下,获得了BSA的校准曲线。总共19种盐[Na _2SO _4,(NH _4)_2SO _4,MgSO _4,LiSO _4,Na _2HPO _4,磷酸钠缓冲液(pH 7.0),NaH _2PO _4,K _2HPO _4,磷酸钾缓冲液(pH 7.0) ,KH _2PO _4,C _6H _8O _7,Na _3C _6H _5O _7,KCHO _2,NaCHO _2,NaCO _3,NaHCO _3,C _2H _4O _2,乙酸钠缓冲液(pH 4.5)和NaC _2H _3O _2]和7聚合物测试[PEG 4000,PEG 8000,PEG 20000,UCON 3900,Ficoll 70000,PES 100000和PVP 40000],并将每条校准曲线与水中BSA的校准曲线进行比较。某些浓度的盐和聚合物对BSA校准曲线有很大影响。碳酸盐是造成最大的盐干扰的原因,而柠檬酸和乙酸即使在最大测试浓度(5 wt%)下也不会产生干扰。在这些聚合物中,UCON产生的干扰最大,而当浓度高达10 wt%时,Ficoll不会产生干扰。结论是,需要在蛋白质定量之前对样品进行方便的稀释,以确保不会受到ATPS相形成成分的明显干扰。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号