首页> 外文期刊>Analytica chimica acta >Coupling sequential injection on-line preconcentration by means of a renewable microcolumn with ion-exchange beads with detection by electrothermal atomic absorption spectrometry - Comparing the performance of eluting the loaded beads with transporti
【24h】

Coupling sequential injection on-line preconcentration by means of a renewable microcolumn with ion-exchange beads with detection by electrothermal atomic absorption spectrometry - Comparing the performance of eluting the loaded beads with transporti

机译:通过可再生微柱与离子交换珠联用的顺序预注入在线预浓缩与电热原子吸收光谱法的检测耦合-比较转运蛋白洗脱负载珠的性能

获取原文
获取原文并翻译 | 示例

摘要

The design of a flow injection/sequential injection (FIA/SIA) on-line preconcentration system incorporating a renewable microcolumn with ion-exchange beads and interfaced with an electrothermal atomic absorption spectrometry (ETAAS) detector is described, and its practical applicability is demonstrated for the determination of trace level concentrations of nickel. The column is loaded with a defined volume of SP Sephadex C-25 cation-exchange resin beads and subsequently exposed to a metered Volume of sample solution. The analyte loaded beads are afterwards eluted with 30 mul of diluted nitric acid, and the eluate transported via air segmentation into the graphite tube. The used beads are then discarded. The ETAAS determination is performed in parallel with the preconcentration process of the ensuing sample. An improved precision of R.S.D. 1.5% is obtained as compared to the 3.4% for the procedure in which the loaded beads are transported directly to the graphite furnace for pyrolysis and atomization, and even improved in comparison to the traditional unidirectional and bidirectional repetitive elution procedures which under comparable conditions yield R.S.D.-values of 5.8 and 4.9%, respectively. The tolerance limits for cations such as Pb(II), Zn(II), Co(II) and Mn(II) were improved up to 10-50-folds, and the linear calibration range extended to comprise 0.02-1.20 mug l(-1). Because of lower operating temperature, the life time of the graphite tube is extended. An enrichment factor of 71.1 and a detection limit of 10.2 ng l(-1) along with a sampling frequency of 12h(-1) were obtained, which are at the same levels as those for the previously described procedure without elution. The present approach was validated by determination of the nickel contents in two certified reference materials, an industrial waste water sample and a human urine sample. (C) 2001 Elsevier Science B.V. All rights reserved. [References: 19]
机译:描述了一种流动注射/顺序注射(FIA / SIA)在线预浓缩系统的设计,该系统结合了带有离子交换珠的可再生微柱并与电热原子吸收光谱仪(ETAAS)检测器连接,并证明了其实际适用性痕量镍浓度的测定。在色谱柱中加载一定体积的SP Sephadex C-25阳离子交换树脂珠,然后将其暴露于定量的样品溶液中。然后,将装有分析物的珠子用30 ul稀硝酸洗脱,并将洗脱液通过空气分段输送到石墨管中。然后丢弃用过的珠子。 ETAAS测定与后续样品的预浓缩过程同时进行。 R.S.D.的改进精度与将装载的珠粒直接运输到石墨炉进行热解和雾化的程序的3.4%相比,获得了1.5%的结果,与传统的单向和双向重复洗脱程序相比,甚至在可比条件下产生RSD的过程中,甚至得到了改善-分别为5.8%和4.9%。阳离子(例如Pb(II),Zn(II),Co(II)和Mn(II))的耐受极限提高了10-50倍,线性校准范围扩大到0.02-1.20杯l( -1)。由于较低的工作温度,延长了石墨管的使用寿命。获得的富集系数为71.1,检测限为10.2 ng l(-1),采样频率为12h(-1),与先前描述的无需洗脱的程序处于相同水平。通过测定两种认证参考材料(工业废水样品和人尿样品)中的镍含量,验证了本方法。 (C)2001 Elsevier Science B.V.保留所有权利。 [参考:19]

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号