首页> 外文期刊>Analytica chimica acta >DETERMINATION OF TRACE METALS IN NATURAL WATERS BY FLAME ATOMIC ABSORPTION SPECTROMETRY FOLLOWING ON-LINE ION-EXCHANGE PRECONCENTRATION
【24h】

DETERMINATION OF TRACE METALS IN NATURAL WATERS BY FLAME ATOMIC ABSORPTION SPECTROMETRY FOLLOWING ON-LINE ION-EXCHANGE PRECONCENTRATION

机译:在线离子交换富集-火焰原子吸收光谱法测定天然水中的痕量金属

获取原文
获取原文并翻译 | 示例
           

摘要

An on-line column preconcentration technique for flow-injection name atomic absorption spectrometry has been developed for the determination of trace metals in natural waters. A minicolumn (20 mm x 2 mm i.d.) filled with a poly(aminophosphonic acid) chelating resin (PAPhA, 20-30 mesh) was used for the preconcentration procedure. For 2 mi samples, detection limits of 0.5, 5.0, 1.5, 1.6, 3.5, 0.6, 3.2, 3.1 and 0.4 mu g l(-1) for Cd, Co, Cr, Cu, Fe, Mn, Ni, ph and Zn, respectively, were achieved. The precision (coefficient of variation) obtained was in the range between 1.1 (Cd and Cu) and 3.3% (Co). Accuracy was also investigated and results agreed with the certified values of the reference material. The sampling rate was 48 h(-1). The method was applied to the determination of trace metals in natural waters samples from Galicia (Spain). [References: 20]
机译:已经开发了一种用于流动注射名称原子吸收光谱的在线柱预富集技术,用于测定天然水中的痕量金属。预填充程序使用装有聚(氨基膦酸)螯合树脂(PAPhA,20-30目)的微型色谱柱(内径20 mm x 2 mm)。对于2 mi样品,Cd,Co,Cr,Cu,Fe,Mn,Ni,ph和Zn的检出限分别为0.5、5.0、1.5、1.6、3.5、0.6、3.2、3.1和0.4 mu gl(-1),分别实现。获得的精度(变异系数)在1.1(Cd和Cu)至3.3%(Co)之间。还对准确性进行了调查,结果与标准物质的认证值相符。采样率为48 h(-1)。该方法用于测定加利西亚(西班牙)天然水样中的痕量金属。 [参考:20]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有