首页> 外文期刊>高等学校化学研究(英文版) >On-line Preconcentration and Determination of Trace Metals in Water Samples by Flow Injection Combined with Flame Atomic Absorption Spectrometry via Calix4arene Carboxylic Acid Packed Micro-column
【24h】

On-line Preconcentration and Determination of Trace Metals in Water Samples by Flow Injection Combined with Flame Atomic Absorption Spectrometry via Calix4arene Carboxylic Acid Packed Micro-column

机译:杯装4芳烃羧酸微柱在线进样富集-流动注射-火焰原子吸收光谱法测定水样中的痕量金属

获取原文
获取原文并翻译 | 示例
       

摘要

An adsorbent calix[4]arene carboxylic acid was employed as the adsorption material for on-line flow in jection(FI) micro-column preconcentration coupled with flame atomic absorption spectrometry(FAAS) determination of trace heavy metals(Cu, Pb, Co, Ni and Cd). Parameters such as the pH, loading time and flow rate of sample, and the concentration, volume and flow rate of eluent were optimized. The enrichment factors are 50.0, 56.5, 11.6, 12.1 and 19.1 for Cu, Pb, Co, Ni, and Cd, respectively, and a sample throughput of 20 h-1 was obtained. The limits of de tection for Cu, Pb, Co, Ni, and Cd were in a range of 1.56―3.91 μg/L, and the relative standard deviations(RSDs) were less than 2.76%(n=7). Furthermore, the proposed method was successfully applied to the determination of Cu, Pb, Co, Ni, and Cd in certified reference materials and various water samples.