首页> 外文期刊>Analytical chemistry >Integrated electrophoresis chips/amperometric detection with sputtered gold working electrodes
【24h】

Integrated electrophoresis chips/amperometric detection with sputtered gold working electrodes

机译:带有电泳金工作电极的集成电泳芯片/安培检测

获取原文
获取原文并翻译 | 示例
       

摘要

An on-chip electrochemical detector for micromachined capillary electrophoresis (CE) systems, based on sputtering a gold working electrode directly onto the capillary outlet, is described. The new on-chip detector preparation requires no microfabrication or alignment procedures nor a decoupling mechanism. The attractive performance of the integrated electrophoresis chips/amperometric detection was demonstrated for the anodic detection of neurotransmitters. The response for dopamine was linear from 20 to 200 mu M, with a LOD of 1.0 mu M and a sensitivity of 52 pA/mu M. Such intimate coupling of capillary electrophoresis chips and electrochemical detection facilitates the realization of complete integrated microanalytical devices. [References: 8]
机译:描述了一种用于微机械毛细管电泳(CE)系统的芯片上电化学检测器,该检测器基于将金工作电极直接溅射到毛细管出口上。新的片上检测器准备工作不需要微加工或对准程序,也不需要去耦机制。集成电泳芯片/安培检测的吸引力表现出了神经递质的阳极检测。多巴胺的响应是线性的,从20到200μM,LOD为1.0μM,灵敏度为52 pA /μM。毛细管电泳芯片和电化学检测的这种紧密耦合有助于实现完整的集成微分析装置。 [参考:8]