首页> 外文期刊>Biomedical Chromatography: An International Journal Devoted to Research in Chromatographic Methodologies and Their Applications in the Biosciences >Development and validation of a sensitive assay for the quantification of imatinib using LC/LC-MS/MS in human whole blood and cell culture.
【24h】

Development and validation of a sensitive assay for the quantification of imatinib using LC/LC-MS/MS in human whole blood and cell culture.

机译:开发和验证用于在人全血和细胞培养物中使用LC / LC-MS / MS定量伊马替尼的灵敏测定方法。

获取原文
获取原文并翻译 | 示例
           

摘要

We developed and validated a semi-automated LC/LC-MS/MS assay for the quantification of imatinib in human whole blood and leukemia cells. After protein precipitation, samples were injected into the HPLC system and trapped onto the enrichment column (flow 5 mL/min); extracts were back-flushed onto the analytical column. Ion transitions [M + H](+) of imatinib (m/z = 494.3 --> 394.3) and its internal standard trazodone (372.5 --> 176.3) were monitored. The range of reliable response was 0.03-75 ng/mL. The inter-day precisions were: 8.4% (0.03 ng/mL), 7.2% (0.1 ng/mL), 6.5% (1 ng/mL), 8.2% (10 ng/mL) and 4.3% (75 ng/mL) with no interference from ion suppression. Autosampler stability was 24 hs and samples were stable over three freeze-thaw cycles. This semi-automated method is simple with only one manual step, uses a commercially available internal standard, and has proven to be robust in larger studies.
机译:我们开发并验证了半自动化LC / LC-MS / MS测定法,用于定量测定人全血和白血病细胞中的伊马替尼。蛋白质沉淀后,将样品注射到HPLC系统中并捕获到富集柱上(流速5 mL / min)。将萃取物反吹到分析柱上。监测了伊马替尼(m / z = 494.3-> 394.3)及其内标曲唑酮(372.5-> 176.3)的离子跃迁[M + H](+)。可靠响应的范围是0.03-75 ng / mL。日间精度为:8.4%(0.03 ng / mL),7.2%(0.1 ng / mL),6.5%(1 ng / mL),8.2%(10 ng / mL)和4.3%(75 ng / mL) ),不受离子抑制的干扰。自动进样器的稳定性为24 h,样品在三个冻融循环中保持稳定。这种半自动化的方法非常简单,只需一个手动步骤,使用了市售的内标,并且在较大的研究中被证明是可靠的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号