A novel microchip capillary electrophoresis system with electrochemical detection, using the replaceable microelectrode, is first reported. This kind of electrode can be fabricated in general laboratories and can be replaced quickly with electrodes of different materials according to the requirements of experiments. The end-column electrochemical detection on microchip CE was achieved by fixing the working electrode (such as carbon fiber, Pt, or An, etc.) through a guide tube on the end of the separation channel. The experiment results indicate that the alignment of the electrode with the channel outlet can be carried out accurately and reproducibly, and therefore, the detection device has low noise and good reproducibility. The detection limit of dopamine is 2.4 x 10(-7) M, which is the lowest result reported so far. The separation and detection of dopamine, 5-hydroxytryptamine and epinephrine using carbon fiber and Pt microdisk electrodes within 50 s was successfully performed. [References: 39]
机译:首次报道了使用可更换微电极的新型带有电化学检测的微芯片毛细管电泳系统。这种电极可以在一般实验室中制造,并可以根据实验要求用不同材料的电极快速更换。通过将工作电极(例如碳纤维,Pt或An等)固定在分离通道末端的导管上,可以在微芯片CE上进行末端柱电化学检测。实验结果表明,该电极与通道出口的对准可以准确,可重复地进行,因此该检测装置噪声低,可重复性好。多巴胺的检出限为2.4 x 10(-7)M,是迄今为止报道的最低结果。成功地在50 s内使用碳纤维和Pt微型电极对多巴胺,5-羟基色胺和肾上腺素进行了分离和检测。 [参考:39]