The matrix effects (MEs) on the quantification of an analyte can be significant and should not be neglected during development and validation of an analytical method. According to this premise, we developed a standardized procedure based on a set of six tests performed on six different sample matrices to detect and characterize the effects of the matrix for single and multiple analytes methods. The link between the matrix effect, recovery, process efficiency, accuracy, precision, and calibration curve was underscored by calculations performed with peak areas, ratios of standard/internal standard peak area, and concentrations. The terms instrumental ME and global ME were introduced, and the term recovery was subdivided for clarity. The test accounts for the presence of ubiquitous and endogenous analytes through background subtraction. The results showed the necessity for using samples with an original concentration in the same range and that the concentration selected for the addition had a definite impact on the results. The use of six-sample matrices provided a standard deviation on the results, and this information could be inserted in a method performance result to show precision. The tool also allows for testing of different analytes/internal standard combinations, which helps with the selection of the association with minimum MEs. A UPLC-MS/MS method for the quantification of several phthalate metabolites in urine was developed and validated with this test. This methodology responds to a scientific need for homogeneity, clarity, and understanding of the results and facilitates the decision-making process while lowering the required costs and time.
机译:分析物的定量的基质效应(MES)可能是显着的,并且在分析方法的开发和验证期间不应被忽略。根据这个前提,我们开发了一种基于在六种不同的样本矩阵上执行的六个测试的标准化过程,以检测和表征矩阵对单一和多个分析物方法的影响。通过使用峰面积,标准/内标峰面积和浓度的峰值面积,峰值,矩阵效应,恢复,处理效率,精度,精度和校准曲线之间的链接。介绍了这个条款,我和全球我被介绍,术语恢复被细分为清晰度。测试占通过背景减法存在普遍存在的和内源性分析物。结果表明,在相同范围内使用原始浓度的样品的必要性,并且为此添加的浓度对结果产生了明确的影响。使用六个样本矩阵提供了对结果的标准偏差,并且可以在方法性能结果中插入该信息以显示精度。该工具还允许测试不同的分析物/内部标准组合,这有助于选择与最小MES的关联。开发并验证了该试验的尿液中几种邻苯二甲酸酯代谢物的UPLC-MS / MS方法。这种方法响应了对结果,清晰度和对结果的理解的科学需求,并促进了决策过程,同时降低了所需的成本和时间。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号