Biomechanical Comparison of Pedicle Screw Augmented with Different Volumes of Polymethylmethacrylate in Osteoporotic and Severely Osteoporotic Synthetic Bone Blocks in Primary Implantation: An Experimental Study

机译:用骨质疏松骨质疏松骨质疏松合成骨嵌段在骨质疏松和严重骨质疏松合成骨嵌段中使用不同体积的椎弓根螺钉的生物力学比较:实验研究

获取原文
获取原文并翻译 | 示例
           

摘要

This study was designed to compare screw stabilities augmented with different volumes of PMMA and analyze relationship between screw stability and volume of PMMA and optimum volume of PMMA in different bone condition. Osteoporotic and severely osteoporotic synthetic bone blocks were divided into groups A0-A5 and B0-B5, respectively. Different volumes of PMMA were injected in groups A0 to A5 and BO to B5. Axial pullout tests were performed and F_(max) was measured. F_(max) in groups A1-A5 were all significantly higher than group A0. Except between groups Al and A2, A3 and A4, and A4 and A5, there were significant differences on F_(max) between any other two groups. F_(max) in groups B1-B5 were all significantly higher than group BO. Except between groups B1 and B2, B2 and B3, and B4 and B5, there were significant differences on F_(max) between any other two groups. There was significantly positive correlation between F_(max) and volume of PMMA in osteoporotic and severely osteoporotic blocks. PMMA can significantly enhance pedicle screw stability in osteoporosis and severe osteoporosis. There were positive correlations between screw stability and volume of PMMA. In this study, injection of 3 mL and 4 mL PMMA was preferred in osteoporotic and severely osteoporotic blocks, respectively.
机译:本研究旨在将螺杆稳定性与不同体积的PMMA进行比较,分析PMMA的螺杆稳定性和体积与不同骨骼条件下的PMMA的最佳体积之间的关系。骨质疏松和严重的骨质疏松合成骨嵌段分别分别分为A0-A5和B0-B5基团。将不同体积的PMMA注射成A0至A5和Bo至B5。执行轴向拉出试验,测量F_(MAX)。 A1-A5组中的F_(MAX)均明显高于A0组。除了AL和A2,A3和A4和A4和A5组之间,在任何其他两组之间的F_(MAX)有显着差异。 B1-B5组中的F_(MAX)均明显高于博组。除了组B1和B2,B2和B3和B4和B5组之间,在任何其他两组之间的F_(MAX)存在显着差异。骨质疏松症和严重骨质疏松障碍中的F_(MAX)和PMMA体积之间存在显着正相关性。 PMMA可以显着提高骨质疏松症和严重骨质疏松症的椎弓根螺杆稳定性。螺杆稳定性和PMMA体积之间存在正相关性。在该研究中,分别在骨质疏松和严重的骨质疏松障碍中注射3mL和4mL PMMA。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号