首页> 外文期刊>Biomedical Chromatography: An International Journal Devoted to Research in Chromatographic Methodologies and Their Applications in the Biosciences >A validated method for analysis of chromium picolinate in nutraceuticals by reversed phase high performance liquid chromatography.
【24h】

A validated method for analysis of chromium picolinate in nutraceuticals by reversed phase high performance liquid chromatography.

机译:一种反相高效液相色谱法分析营养食品中吡啶甲酸铬的有效方法。

获取原文
获取原文并翻译 | 示例
       

摘要

A validated high performance liquid chromatographic method was developed for the determination of chromium picolinate in pharmaceutical dosage forms. The analysis was performed at room temperature using a reversed-phase Supelcosil LC-18 (250 x 4.6 mm, 5 microm) column. The mobile phase consisted of acetonitrile:water (40:60 v/v) at a fl ow rate of 0.8 mL/min. The UV-detector was set at 264 nm. The developed method showed a good linear relationship in the concentration range from 0.125 to 12.5 microg/mL with a correlation coefficient from 0.999. The limit of detection and limit of quanti fi cation were 0.091 and 0.181 microg/mL, respectively.
机译:建立了一种经验证的高效液相色谱方法,用于测定药物剂型中的吡啶甲酸铬。使用反相Supelcosil LC-18(250 x 4.6 mm,5微米)色谱柱在室温下进行分析。流动相由流速为0.8 mL / min的乙腈:水(40:60 v / v)组成。紫外线检测器设定为264nm。所开发的方法在0.125至12.5 microg / mL的浓度范围内表现出良好的线性关系,相关系数为0.999。检测限和定量限分别为0.091和0.181 microg / mL。