首页> 外文期刊>Anaesthesia: Journal of the Association of Anaesthetists of Great Britain and Ireland >Unalarmed interruption of continuous drug infusion by Perfusor Space GC syringe pumps
【24h】

Unalarmed interruption of continuous drug infusion by Perfusor Space GC syringe pumps

机译:Perfusor Space GC注射泵无误中断连续药物输注

获取原文
获取原文并翻译 | 示例
           

摘要

We wish to report two cases of unalarmed syringe pump malfunction (Perfusor Space GC, B. Braun, Melsungen, Germany) that could have resulted in unintended intra-operative awareness.After uneventful testing of all devices, intravenous anaesthesia was induced in the first patient and subsequently maintained by propofol infusion at 6 mg.kg~(-1).h~(-1). Twenty-five minutes after induction, the patient's heart rate and blood pressure increased without surgical stimulus, so the depth of anaesthesia was increased by administering 0.3 mug.kg~(-1) sufentanil and repeated boluses of propofol via the syringe pump's bolus function (2 mg.kg~(-1) over 10 min), with the propofol flow rate increased from 6 to 12 mg.kg~(-1).h~(-1). Despite this, the patient moved his extremities. Suspecting a syringe pump dysfunction, a bolus of 1.4-mg.kg~(-1) propofol was administered manually, restoring clinical anaesthesia. The syringe pump was exchanged and surgery was accomplished without problems.
机译:我们希望报告2例未发生警报的注射泵故障(Perfusor Space GC,B.Braun,Melsungen,Germany),这可能会导致意想不到的术中意识。在对所有设备进行无损检测之后,第一例患者进行了静脉麻醉然后通过异丙酚输注维持在6 mg.kg〜(-1).h〜(-1)。诱导后25分钟,患者的心率和血压升高而无手术刺激,因此通过注射0.3杯·kg〜(-1)的舒芬太尼并通过注射泵的推注功能反复推注异丙酚来增加麻醉深度。在10分钟内2 mg.kg〜(-1)),异丙酚流速从6增加到12 mg.kg〜(-1).h〜(-1)。尽管如此,病人还是移动了四肢。怀疑注射泵功能异常,手动给予1.4 mg.kg〜(-1)异丙酚大剂量,恢复了临床麻醉效果。更换了注射泵,手术顺利完成。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号