Checking anaesthetic equipment 2012: Association of anaesthetists of Great Britain and Ireland

机译:检查麻醉设备2012:大不列颠及爱尔兰麻醉师协会

获取原文
获取原文并翻译 | 示例
           

摘要

Summary A pre-use check to ensure the correct functioning of anaesthetic equipment is essential to patient safety. The anaesthetist has a primary responsibility to understand the function of the anaesthetic equipment and to check it before use. Anaesthetists must not use equipment unless they have been trained to use it and are competent to do so. A self-inflating bag must be immediately available in any location where anaesthesia may be given. A two-bag test should be performed after the breathing system, vaporisers and ventilator have been checked individually. A record should be kept with the anaesthetic machine that these checks have been done. The 'first user' check after servicing is especially important and must be recorded.
机译:小结进行使用前检查,以确保麻醉设备的正常运行对于患者安全至关重要。麻醉师的主要职责是了解麻醉设备的功能并在使用前进行检查。麻醉师不得使用设备,除非他们经过培训可以使用它。自充气袋必须在可能进行麻醉的任何位置立即可用。分别检查了呼吸系统,气化器和呼吸机后,应进行两袋测试。麻醉机应保留已完成这些检查的记录。维修后的“第一用户”检查尤为重要,必须加以记录。