【24h】

Inaccurate colour coding of medical gas cylinders.

机译:医用气瓶的颜色编码不正确。

获取原文
获取原文并翻译 | 示例
           

摘要

We would like to highlight a further case of inaccurate labelling of medical gas cylinders [1]. The anaesthetic machine in our Magnetic Resonance Imaging unit is supplied with cylinders of oxygen and nitrous oxide from BOC Medical (Worsley, Manchester, UK). These cylinders are both size AZ, suitable for use within the Magnetic Resonance Imaging environment. Confusingly, both oxygen and nitrous oxide are given a green body colour. The shoulder colour (white and blue, respectively) and collar labels conform to European regulations [2].
机译:我们想强调一下医用气瓶标签不准确的另一种情况[1]。我们的磁共振成像装置中的麻醉机配有BOC Medical(英国曼彻斯特沃斯利)的氧气和一氧化二氮气瓶。这些圆柱体均为AZ尺寸,适合在磁共振成像环境中使用。令人困惑的是,氧气和一氧化二氮均呈绿色。肩部颜色(分别为白色和蓝色)和项圈标签符合欧洲法规[2]。