A design fault of the Drager Cato anaesthesia workstation.

机译:Drager Cato麻醉工作站的设计故障。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Recently there have been reports of some newer anaesthetic machines not alarming when there was discontinuation of the breathing circuit13. The breathing circuit was opened to air, resulting in progressive decline in the inspired concentration of anaesthetic agents. This places patients at risk of hypoxia and awareness. I would like to report another fault with the Drager Cato anaesthesia machine (Drager Medical, Germany). It appears that when there is a disconnection at the common gas outlet during positive pressure ventilation, whether in volume control or pressure control modes, the machine does not alarm.
机译:最近有报道说,一些新的麻醉机在呼吸回路中断时不会发出警报13。呼吸回路向空中开放,导致吸入的麻醉药浓度逐渐下降。这使患者处于缺氧和意识状态的风险。我想报告Drager Cato麻醉机(德国Drager Medical)的另一个故障。在正压通风过程中,当公用气体出口断开时,无论是在音量控制还是在压力控制模式下,机器都不会发出警报。