Comparison of stainless steel and titanium alloy orthodontic miniscrew implants: A mechanical and histologic analysis

机译:不锈钢和钛合金正畸微螺钉种植体的比较:力学和组织学分析

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The detailed mechanical and histologic properties of stainless steel miniscrew implants used for temporary orthodontic anchorage have not been assessed. Thus, the purpose of this study was to compare them with identically sized titanium alloy miniscrew implants. Methods: Forty-eight stainless steel and 48 titanium alloy miniscrew implants were inserted into the tibias of 12 rabbits. Insertion torque and primary stability were recorded. One hundred grams of tensile force was applied between half of the implants in each group, resulting in 4 subgroups of 24 specimens each. Fluorochrome labeling was administered at weeks 4 and 5. When the rabbits were euthanized at 6 weeks, stability and removal torque were measured in half (ie, 12 specimens) of each of the 4 subgroups. Microdamage burden and bone-to-implant contact ratio were quantified in the other 12 specimens in each subgroup. Mixed model analysis of variance was used for statistical analysis. Results: All implants were stable at insertion and after 6 weeks. The only significant difference was the higher (9%) insertion torque for stainless steel. No significant differences were found between stainless steel and titanium alloy miniscrew implants in microdamage burden and bone-to-implant contact regardless of loading status. Conclusions: Stainless steel and titanium alloy miniscrew implants provide the same mechanical stability and similar histologic responses, suggesting that both are suitable for immediate orthodontic clinical loads.
机译:未对用于临时正畸固定的不锈钢微螺钉植入物的详细力学和组织学特性进行评估。因此,本研究的目的是将它们与相同尺寸的钛合金微螺钉植入物进行比较。方法:将48只不锈钢和48只钛合金微螺钉植入物插入12只兔子的胫骨中。记录插入扭矩和一次稳定性。在每组一半的植入物之间施加一百克的拉力,形成4个亚组,每组24个标本。在第4和第5周进行荧光染料标记。在第6周对安乐死进行兔子处死后,在4个亚组的一半(即12个样本)中测量稳定性和去除扭矩。在每个亚组的其他12个样本中,对微损伤负担和骨与植入物的接触比进行了定量。方差的混合模型分析用于统计分析。结果:所有植入物在插入时和6周后均稳定。唯一的显着差异是不锈钢的插入扭矩更高(9%)。在不锈钢和钛合金微型螺钉植入物之间,无论载荷状态如何,其微损伤负荷和骨与植入物之间的接触均无显着差异。结论:不锈钢和钛合金微螺钉植入物提供相同的机械稳定性和相似的组织学反应,表明两者均适合于即时正畸临床负荷。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号