Scheimpflug corneal power measurements for intraocular lens power calculation in cataract surgery

机译:白内障手术中Scheimpflug角膜屈光度测量用于人工晶状体计算

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To compare the keratometric (K) readings from the Pentacam-HR (High Resolution) unit with the automated K values from the IOLMaster keratometer (KIOLM), and to evaluate them in the commonly used intraocular lens (IOL) power calculation formulas for routine cataract surgery. Design: Prospective, comparative observational study. Methods: setting: Private practice, Lynwood, California. study population: Fifty cataractous eyes scheduled for surgery between July and August 2012. observation procedure: The K readings from the Pentacam-HR unit taken at the 2-, 3-, 4-, and 5-mm zones and the 2-, 3-, 4-, and 5-mm rings, respectively, from 3 different maps: sagittal corneal front (KF), true net power (KTNP), and total refractive power (KRP) are compared with KIOLM. IOL power calculations were performed with each of the 25 sets of K readings. main outcome measures: The IOL prediction median absolute error (MedAE) obtained with each measurement. Results: KF averaged 0.03-0.13 diopter (D) higher than KIOLM (P .05), KTNP averaged 1.16-1.21 D lower than KIOLM (P .001), and KRP averaged 0.23-0.72 D lower than KIOLM (P .001), with large variations in the measurements. The MedAE obtained with the different Pentacam K readings ranged from 0.44-0.64 D vs 0.52 D obtained with KIOLM (P .05). MedAE was lower in all categories when the pupil was 3 mm or smaller. Conclusion: The Pentacam K F values were the closest to KIOLM and the KF readings from the 2-mm ring yielded the best results for IOL power calculation.
机译:目的:将Pentacam-HR(高分辨率)装置的角膜曲率(K)读数与IOLMaster角膜曲率计(KIOLM)的自动K值进行比较,并在常用的人工晶状体(IOL)屈光力计算公式中进行评估常规白内障手术。设计:前瞻性,比较性观察研究。方法:环境:私人诊所,加利福尼亚州林伍德。研究人群:计划在2012年7月至2012年8月之间进行手术的白内障眼睛为50只。观察程序:从Pentacam-HR装置的2-,3-,4-和5-mm区域以及2-,3区域获取K读数将3个不同图中的-,4和5毫米环分别与KIOLM进行比较:矢状角膜前缘(KF),真实净屈光力(KTNP)和总屈光力(KRP)。用25组K读数中的每组进行IOL功率计算。主要结果度量:每次测量获得的IOL预测中位数绝对误差(MedAE)。结果:KF平均比KIOLM高0.03-0.13 D(P> .05),KTNP平均比KIOLM低1.16-1.21 D(P> .001),KRP平均比KIOLM低0.23-0.72 D(P> (.001),测量值会有很大差异。使用不同的Pentacam K读数获得的MedAE范围为0.44-0.64 D,而使用KIOLM获得的读数为0.52 D(P> .05)。当瞳孔小于等于3 mm时,MedAE在所有类别中均较低。结论:Pentacam K F值最接近KIOLM,并且2 mm环的KF读数为IOL功率计算提供了最佳结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号