A method for urinary iodine diagnostics using a portable energy-dispersion X-ray fluorescent (EDXRF) spectrometer is proposed. The principle of the method consists in excitation of the sample atoms fluorescence by the energy spectrum intentionally formed from the spectrum of an X-ray tube using an optimized X-ray scheme. The optimization by the criterion of the minimum detection limit for L-series iodine fluorescence lines included the calculation of optimal atomic numbers for materials of a re-radiator and a filter, and their thicknesses, the collimation system parameters, and an X-ray tube voltage. Experimentally achieved detection limits were 45 and 75 mu g/L by I-L alpha and I-L beta lines, respectively, at the theoretically extreme value 30 mu g/L. This sensitivity is found to be sufficient for urinary iodine diagnostics in the range from 50 to 200 mu g/L. The results obtained from different patients have shown the satisfactory convergence for the iodine concentration determination by L alpha and I-L beta fluorescence lines. The simple sample preparation procedure and comparably small sizes of the apparatus allows rapid researches directly in clinical settings.
机译:提出了一种使用便携式能量 - 色散X射线荧光(EDXRF)光谱仪的尿碘诊断方法。该方法的原理在于通过使用优化的X射线方案特意由X射线管的光谱特征形成的能谱激发样品原子荧光。通过L系列碘荧光线的最小检测限的标准的优化包括用于重新辐射器和过滤器的材料的最佳原子编号,以及它们的厚度,准直系统参数和X射线管电压。通过I-Lα和I-Lβ线分别在实验上实现的检测限为45和75μg/ L,在理论上极值30μg/升。发现这种敏感性足以用于尿碘诊断的范围为50至200μg/升。由不同患者获得的结果表明,通过Lα和I-Lβ荧光线对碘浓度测定的令人满意的会聚。简单的样品制备程序和相当小的装置允许直接在临床环境中进行快速研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号