首页> 外文期刊>Diabetes technology & therapeutics >Glutest Neo Super--a new handheld blood glucose meter-corrects for the effects of the hematocrit values in both hematocrit-adjusted samples and samples obtained from anemic patients.
【24h】

Glutest Neo Super--a new handheld blood glucose meter-corrects for the effects of the hematocrit values in both hematocrit-adjusted samples and samples obtained from anemic patients.

机译:最佳Neo Super - 一种新的手持式血糖仪 - 校正血细胞比容调整样品中的血细胞比容值和从贫血患者获得的样品的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Handheld blood glucose (BG) meters are convenient tools that are widely used to measure the BG levels. However, the hematocrit (Hct) value has been identified as a confounding factor for accurate BG measurement. Some BG meters are equipped with an Hct-correcting feature, whose effectiveness has been tested previously. Nevertheless, the measurements yielded by many BG meters are confounded by the Hct values. Recently, a new BG meter equipped with an Hct-correcting feature has been developed; however, its effectiveness has not yet been confirmed. STUDY DESIGN: Venous blood samples were collected from two healthy volunteers, and the Hct values in the samples were adjusted to approximately 0%, 10%, 20%, 30%, 40%, 50%, and 60%. Further, venous blood samples were collected from 10 anemic patients (Hct <40%). The whole BG (WBG) levels in the samples were measured using two devices-the new BG meter (Glutest Neo Super [Sanwa Kagaku Kenkyusho Co. Ltd., Nagoya, Japan]) and a standard BG meter (OneTouch Ultra [Life Scan Inc., Milpitas, CA]). For reference, plasma glucose (PG) levels were measured using a machine at our hospital laboratory (GA08 [A&T Co., Kanagawa, Japan]). The bias in the measurements was calculated as follows: bias = ([WBG - PG]/PG) x 100. Further, the correlation between the Hct values and the bias was assessed by performing linear regression analysis. RESULTS: In both the Hct-adjusted samples and the samples obtained from anemic patients, the WBG levels measured using Glutest Neo Super were minimally affected by the Hct values, while those measured using OneTouch Ultra were affected by the Hct values to a statistically significant extent. CONCLUSIONS: The Hct-correcting feature of the new BG meter Glutest Neo Super was effective. The use of this new device for BG measurements may lead to more appropriate treatment selection.
机译:背景:手持式血糖(BG)计是广泛用于测量BG水平的便利工具。然而,血细胞比容(HCT)值已被鉴定为精确BG测量的混淆因素。一些BG仪表配备了HCT校正功能,其效率先前已经过测试。然而,许多BG仪表产生的测量由HCT值混淆。最近,已经开发出一个配备HCT校正功能的新BG仪表;但是,它的有效性尚未得到证实。研究设计:从两个健康的志愿者收集静脉血样,并将样品中的HCT值调节至约0%,10%,20%,30%,40%,50%和60%。此外,从10名贫血患者(HCT <40%)收集静脉血样。样品中的整个BG(WBG)水平使用两种器件 - 新的BG仪表(Glutest Neo Super [Sanwa Kagaku Kenkyusho Co.Tanks,Nagoya,Japan])和标准的BG仪表(OneTouch Ultra [Life Scan Inc 。,米尔皮塔斯,加利福尼亚州)。参考,使用医院实验室(GA08 [A&T Co.,Kanagawa,Japan])的机器测量血浆葡萄糖(PG)水平。测量中的偏差计算如下:偏压=([WBG-PG] / pg)×100.此外,通过执行线性回归分析来评估HCT值与偏差之间的相关性。结果:在HCT调整后的样品和从贫血患者获得的样品中,使用最低型高超的WBG水平受到HCT值的最小影响,而使用OneTouch Ultra测量的那些将受到HCT值的影响,以统计学显着影响。结论:新型BG仪表最低型新超级超级的HCT校正特征是有效的。这种用于BG测量的新设备可能导致更合适的治疗选择。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号