AbstractA 58-year-old man undergoing cardiac resynchronization therapy with a defibrillator in his right subcostal area fell from his bed, leading to fracture of the right clavicle. Serial radiographs showed dislocation of the distal clavicle 2 months after the initial fracture. Lead parameters dramatically changed after dislocation of the distal clavicle. Radiography indicated that the device leads seemed to be compressed by the distal clavicle in certain positions of the right upper limb. It was likely that various movements of the right upper limb during his daily life insidiously damaged the device leads, leading to the lead breakages.]]>
机译:<![CDATA [ 抽象 一个58岁的男子在他的右侧底板区域中占用了除颤器的心脏重新同步疗法,导致右锁骨骨折。序列射线照片显示初始骨折后2个月的远端锁骨​​脱位。在远端锁定后脱位后,铅参数显着改变。射线照相表明,器件引线似乎被右上肢的某些位置中的远端锁骨​​压缩。在他的日常生活中,右上肢的各种运动意外地损坏了设备的导线,导致铅断裂。 ] ]