Method validation studies characterize the performance of new laboratory methods relative to established methods using quality guidelines in order to define the new method's performance characteristics and to identify differences that could influence data interpretation. We investigated the performance of an in-clinic dry chemistry analyzer (Catalyst One, IDEXX) for measuring 19 routine plasma biochemistry analytes in dogs, cats, and horses. We analyzed 2 levels of quality control material (QCM) in duplicate twice daily for 5 d to determine the coefficient of variation (CV), percent bias, observed total error (TEobs), and sigma metric (sigma) for each analyte at each level of QCM. We analyzed 82 canine, equine, and feline plasma samples with the in-clinic dry chemistry analyzer and a reference wet chemistry analyzer, and results were compared using correlation coefficients, Deming regression, and Bland-Altman analyses. CVs were &5% for 16 analytes and & 5% for 3 analytes. TEobs was less than allowable total error (TEa) for 9 analytes, and exceeded TEa for 10 analytes. Sigma metrics were &4 at both levels of QCM for 5 analytes, and at one level of QCM for 5 analytes; sigma metrics were &3 or could not be calculated at the remaining analyte concentrations. All analytes, except glucose, showed various magnitudes of bias compared to the wet chemistry analyzer. Based on these results, we recommend statistical (5 analytes) and non-statistical (14 analytes) QC measures and analyzer-specific reference intervals.
机译:方法验证研究表征了使用质量指南的新实验室方法的性能,以定义新方法的性能特征,并识别可能影响数据解释的差异。我们调查了临床干燥化学分析仪(催化剂1,IDEXX)的表现,用于测量狗,猫和马的常规血浆生物化学分析物。我们分析了2层水平的质量控制材料(QCM)两次每天两次,5 d确定每个级别的每个分析物的变异系数(CV),百分比,观察到的总误差(Teobs)和Sigma度量(Sigma)。 QCM。我们分析了82只犬,马蹄和翅膀的血浆样品与临床干燥化学分析仪和参考湿化学分析仪,并使用相关系数,脱亮回归和平坦altman分析进行比较结果。 CVS是& LT; 16%的分析物和& 3分析物5%。 Teobs少于允许的9分析物的总误差(茶),超过10种分析物的茶。 Sigma度量均为& 4在QCM的含量为5分析物,并且在5个分析物的QCM的一级; Sigma度量是& LT; 3或不能以剩余的分析物浓度计算。除葡萄糖外,所有分析物均显示与湿化学分析仪相比的各种偏差幅度。基于这些结果,我们推荐统计(5分析物)和非统计(14分析物)QC措施和分析仪特定的参考间隔。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号