A novel method for comparing radiation dose and image quality, between and within different x-ray units in a series of hospitals

机译:一种比较辐射剂量和图像质量的新方法,在一系列医院中的不同X射线单元之间和内部

获取原文
获取原文并翻译 | 示例
       

摘要

Objectives: To develop a novel method for comparing radiation dose and image quality (IQ) to evaluate adult chest x-ray (CXR) imaging among several hospitals. Methods: CDRAD 2.0 phantom was used to acquire images in eight hospitals (17 digital x-ray units) using local adult CXR protocols. IQ was represented by image quality figure inverse (IQF(inv)), measured using CDRAD analyser software. Signal to noise ratio, contrast to noise ratio and conspicuity index were calculated as additional measures of IQ. Incident air kerma (IAK) was measured using a solid-state dosimeter for each acquisition. Figure of merit (FOM) was calculated to provide a single estimation of IQ and radiation dose. Results: IQ, radiation dose and FOM varied considerably between hospitals and x-ray units. For IQF(inv), the mean (range) between and within the hospitals were 1.42 (0.83-2.18) and 1.87 (1.52-2.18), respectively. For IAK, the mean (range) between and within the hospitals were 93.56 (17.26-239.15) mu Gy and 180.85 (122.58-239.15) mu Gy, respectively. For FOM, the mean (range) between and within hospitals were 0.05 (0.01-0.14) and 0.03 (0.02-0.05), respectively. Conclusions: The suggested method for comparing IQ and dose using FOM concept along with the new proposed FOM, is a valid, reliable and effective approach for monitoring and comparing IQ and dose between and within hospitals. It is also can be beneficial for the optimisation of x-ray units in clinical practice. Further optimisation for the hospitals/x-ray units with low FOM are required to minimise radiation dose without degrading IQ.
机译:目的:开发一种用于比较辐射剂量和图像质量(IQ)的新方法,以评估几家医院之间的成人胸部X射线(CXR)成像。方法:使用CDRAD 2.0 PHANTOM使用当地成人CXR协议来获取八个医院(17个数字X射线单元)中的图像。 IQ由使用CDRAD Analyzer软件测量的图像质量数字逆(IQF(INV))表示。信噪比,与噪声比对比度和链接率指数的对比度被计算为IQ的额外测量。使用固态剂量计测量入射空气Kerma(IAK),每次采集测量。计算优异(FOM)的数字,以提供IQ和辐射剂量的单一估计。结果:IQ,辐射剂量和FOM在医院和X射线单元之间变化很大。对于IQF(INV),医院之间的平均值(范围)分别为1.42(0.83-2.18)和1.87(1.52-2.18)。对于IAK,医院之间的平均(范围)为93.56(17.26-239.15)MU GY和180.85(122.58-239.15)MU GY。对于FOM,医院之间的平均值(范围)分别为0.05(0.01-0.14)和0.03(0.02-0.05)。结论:使用FOM概念与新拟议FOM进行比较和剂量比较IQ和剂量的建议方法,是一种有效,可靠且有效的方法,用于监测和比较医院之间的IQ和剂量。它还可以有利于在临床实践中优化X射线单元。需要具有低FOM的医院/ X射线单元的进一步优化,以尽量减少辐射剂量而不会降低IQ。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号